Peltier Measurements below 4 K
Author:
- 1 May 1969
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 40 (5) , 648-653
Abstract
Accurate low temperature Seebeck measurements are difficult since they require the detection of very small emf's. An alternative Peltier measurement, which provides equivalent information, is described, and satisfactory experimental results are presented on materials with Seebeck coefficients as low as 2×10−8 V/K. The technique has the capability of measuring a Seebeck coefficient of 10−9 V/K with 2% precision and should be particularly useful for use on high resistivity alloys and for measurements in the presence of a magnetic field.Keywords
This publication has 16 references indexed in Scilit:
- Thomson's dilemmaPhysics Today, 1967
- Thermoelectric power of silver alloys at very low temperaturesPhilosophical Magazine, 1967
- THE ISOTHERMAL PELTIER COEFFICIENT AND PHONON-DRAG IN COPPERCanadian Journal of Physics, 1963
- Electron Transport Phenomena in Zinc at Liquid-Helium TemperaturesPhysical Review B, 1963
- Experimental Verification of the Kelvin Relation of Thermoelectricity in a Magnetic FieldPhysical Review B, 1963
- A sensitive superconducting “chopper” amplifierPhysica, 1957
- The Kelvin relations in thermo-electricityProceedings of the Royal Society of London. A. Mathematical and Physical Sciences, 1956
- Irreversible Thermodynamics of ThermoelectricityReviews of Modern Physics, 1954
- A superconducting galvanometerMathematical Proceedings of the Cambridge Philosophical Society, 1952
- The direct measurement of the Peltier coefficientProceedings of the Physical Society (1926-1948), 1935