Validation of a locally designed computed tomography dose phantom: comparative assessment against a standard acrylic phantom for dose accuracy and consistency in edo and delta states, nigeria
Open Access
- 1 July 2025
- journal article
- Published by HM Publishers in Journal of Nursing Research,Patient Safety and Practise
- Vol. 5 (52) , 1-13
Abstract
This study compares CT dose measurements specifically CTDIvol and DLP between standard phantoms (SP) and locally constructed phantoms (CP) for head and body scans across four CT centres. Locally built phantoms, made from bent PMMA sheets and filled with water, were tested alongside standard phantoms using ionization chambers. Results show minor differences in dose estimates, with head CT showing slightly higher deviations (up to 7.18%) compared to body CT (up to 3.25%). Console-displayed values followed similar trends, with some larger variations noted, especially for head CT at one centre (up to 52.11% deviation). However, ANOVA analysis revealed no statistically significant differences between SP and CP dose values. All results were within the ±20% uncertainty range recommended by IAEA and ACR, suggesting that locally constructed phantoms are suitable alternatives for CT dose assessments in clinical practice.Keywords
This publication has 20 references indexed in Scilit:
- CHARACTERIZATION OF RADIOACTIVITY FROM PRIMORDIAL RADIONUCLIDES IN THE SOIL OF IKA SOUTH LOCAL GOVERNMENT AREA OF DELTA STATEFUDMA JOURNAL OF SCIENCES, 2022
- Mechanism of the radiation-induced transformations of fluoroform in solid noble gas matrixesRadiation Physics and Chemistry, 2017
- U.S. Diagnostic Reference Levels and Achievable Doses for 10 Adult CT ExaminationsRadiology, 2017
- Radiological Evaluation of Bowel IschemiaRadiologic Clinics of North America, 2015
- Determination of dose delivery accuracy in CT examinationsJournal of Radiation Research and Applied Sciences, 2015
- Emerging clinical applications of computed tomographyMedical Devices: Evidence and Research, 2015
- NATIONAL DIAGNOSTIC REFERENCE LEVEL INITIATIVE FOR COMPUTED TOMOGRAPHY EXAMINATIONS IN KENYARadiation Protection Dosimetry, 2015
- Measurements of the dose delivered during CT exams using AAPM Task Group Report No. 111Journal of Applied Clinical Medical Physics, 2012
- Aspirin Plus Clopidogrel and Risk of Infection After Coronary Artery Bypass SurgeryArchives of internal medicine (1960), 2009
- Computed Tomography — An Increasing Source of Radiation ExposureThe New England journal of medicine, 2007