Fixed gantry tomosynthesis system for radiation therapy image guidance based on a multiple source x‐ray tube with carbon nanotube cathodes
- 13 April 2009
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 36 (5) , 1624-1636
Abstract
The authors present the design and simulation of an imaging system that employs a compact multiple source x‐ray tube to produce a tomosynthesis image from a set of projections obtained at a single tube position. The electron sources within the tube are realized using cold cathode carbon nanotube technology. The primary intended application is tomosynthesis‐based 3D image guidance during external beam radiation therapy. The tube, which is attached to the gantry of a medical linear accelerator (linac) immediately below the multileaf collimator, operates within the voltage range of and contains a total of 52 sources that are arranged in a rectilinear array. This configuration allows for the acquisition of tomographic projections from multiple angles without any need to rotate the linac gantry. The x‐ray images are captured by the same amorphous silicon flat panel detector employed for portal imaging on contemporary linacs. The field of view (FOV) of the system corresponds to that part of the volume that is sampled by rays from all sources. The present tube and detector configuration provides an FOV in the plane of the linac isocenter when the imaging detector is placed from the isocenter. Since this tomosynthesis application utilizes the extremities of the detector to record image detail relating to structures near the isocenter, simultaneous treatment and imaging is possible for most clinical cases, where the treated target is a small region close to the linac isocenter. The tomosynthesis images are reconstructed using the simultaneous iterative reconstruction technique, which is accelerated using a graphic processing unit. The authors present details of the system design as well as simulated performance of the imaging system based on reprojections of patient CT images.Keywords
This publication has 28 references indexed in Scilit:
- Low dose megavoltage cone beam computed tomography with an unflattened 4 MV beam from a carbon targetMedical Physics, 2008
- Correlation between dosimetric effect and intrafraction motion during prostate treatments delivered with helical tomotherapyPhysics in Medicine & Biology, 2008
- Megavoltage cone beam digital tomosynthesis (MV-CBDT) for image-guided radiotherapy: a clinical investigational systemPhysics in Medicine & Biology, 2008
- Developments in electronic portal imaging systemsBritish Journal of Radiology, 2006
- Multiplexing radiography using a carbon nanotube based x-ray sourceApplied Physics Letters, 2006
- Stationary scanning x-ray source based on carbon nanotube field emittersApplied Physics Letters, 2005
- Carbon nanotubes as electron source in an x-ray tubeApplied Physics Letters, 2001
- Field-emitting structures intended for a miniature X-ray sourceSensors and Actuators A: Physical, 2000
- Helical microtubules of graphitic carbonNature, 1991
- Simultaneous Algebraic Reconstruction Technique (SART): A superior implementation of the ART algorithmUltrasonic Imaging, 1984