Glutest Neo Super—a New Handheld Blood Glucose Meter—Corrects for the Effects of the Hematocrit Values in Both Hematocrit-Adjusted Samples and Samples Obtained from Anemic Patients
Open Access
- 1 June 2009
- journal article
- research article
- Published by SAGE Publications in Diabetes Technology & Therapeutics
- Vol. 11 (6) , 369-372
Abstract
Background: Handheld blood glucose (BG) meters are convenient tools that are widely used to measure the BG levels. However, the hematocrit (Hct) value has been identified as a confounding factor for accurate BG measurement. Some BG meters are equipped with an Hct-correcting feature, whose effectiveness has been tested previously. Nevertheless, the measurements yielded by many BG meters are confounded by the Hct values. Recently, a new BG meter equipped with an Hct-correcting feature has been developed; however, its effectiveness has not yet been confirmed. Study Design: Venous blood samples were collected from two healthy volunteers, and the Hct values in the samples were adjusted to approximately 0%, 10%, 20%, 30%, 40%, 50%, and 60%. Further, venous blood samples were collected from 10 anemic patients (Hct ® Super [Sanwa Kagaku Kenkyusho Co. Ltd., Nagoya, Japan]) and a standard BG meter (OneTouch ® Ultra ® [Life Scan Inc., Milpitas, CA]). For reference, plasma glucose (PG) levels were measured using a machine at our hospital laboratory (GA08 [A&T Co., Kanagawa, Japan]). The bias in the measurements was calculated as follows: bias = ([WBG − PG]/PG) × 100. Further, the correlation between the Hct values and the bias was assessed by performing linear regression analysis. Results: In both the Hct-adjusted samples and the samples obtained from anemic patients, the WBG levels measured using Glutest Neo Super were minimally affected by the Hct values, while those measured using OneTouch Ultra were affected by the Hct values to a statistically significant extent. Conclusions: The Hct-correcting feature of the new BG meter Glutest Neo Super was effective. The use of this new device for BG measurements may lead to more appropriate treatment selection.Keywords
This publication has 10 references indexed in Scilit:
- The Technology Behind Glucose Meters: Test StripsDiabetes Technology & Therapeutics, 2008
- Evaluation of the Impact of Hematocrit and Other Interference on the Accuracy of Hospital-Based Glucose MetersDiabetes Technology & Therapeutics, 2008
- Error Rates Resulting From Anemia can be Corrected in Multiple Commonly Used Point-of-Care GlucometersJournal of Trauma: Injury, Infection & Critical Care, 2008
- Screening for hypoglycemia at the bedside in the neonatal intensive care unit (NICU) with the Abbott PCx glucose meterBMC Pediatrics, 2006
- Accuracy evaluation of a new glucometer with automated hematocrit measurement and correctionClinica Chimica Acta, 2005
- Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care TestingArchives of Pathology & Laboratory Medicine, 2000
- The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocritEarly Human Development, 2000
- Use of arterial blood with bedside glucose reflectance meters in an intensive care unitCritical Care Medicine, 1994
- Screening for hypoglycemia with plasma in neonatal blood of high hematocrit valueCritical Care Medicine, 1989
- Importance of the Hematocrit in Interpretation of Blood SugarDiabetes, 1965