Biomechanical performance of bicortical versus pericortical bone trajectory (CBT) pedicle screws
- 31 May 2021
- journal article
- research article
- Published by Springer Nature in European Spine Journal
- Vol. 30 (8) , 2292-2300
Abstract
Purpose The cortical bone trajectory (CBT) is an alternative to the traditional pedicle screw trajectory (TT) in posterior spinal instrumentation, enhancing screw contact with cortical bone and therefore increasing fixation strength. Additional to the trajectory, insertion depth (pericortical vs. bicortical placement) could be a relevant factor affecting the fixation strength. However, the potential biomechanical benefit of a bicortical placement of CBT screws is unknown. Therefore, the aim of this study was to quantify the fixation strength of pericortical- versus bicortical-CBT (pCBT versus bCBT) screws in a randomized cadaveric study. Methods Pedicle screws were either placed pericortical or bicortical with a CBT in 20 lumbar vertebrae (2 × 20 instrumented pedicles) from four human spine cadavers by using patient-specific templates. Instrumented specimens underwent physiological cyclic loading testing (1′800′000 cycles, 10 Hz), including shear and tension loads as well as bending moments. Translational and angular displacements of the screws were quantified and compared between the two techniques. Results There was a slight decrease in translational (0.2 mm ± 0.09 vs. 0.24 mm ± 0.11) and angular displacements (0.06° ± 0.05 vs. 0.13° ± 0.11) of bCBT screws when compared with pCBT screws after 1′800′000 cycles. However, the results were non-significant (p > 0.05). Conclusion The authors do not recommend placing CBT screws bicortically, as no relevant biomechanical advantage is gained while the potential risk for iatrogenic injury to structures anterior to the spine is increased.Keywords
Funding Information
- MySpine, Medacta SA International, Switzerland (Vertebra-specific drill guides and the pedicle screws)
This publication has 41 references indexed in Scilit:
- Simultaneous screening for osteoporosis at CT colonography: Bone mineral density assessment using MDCT attenuation techniques compared with the DXA reference standardJournal of Bone and Mineral Research, 2011
- Instrumentation of the osteoporotic spine: biomechanical and clinical considerationsThe Spine Journal, 2011
- Descending Aortic Injury by a Thoracic Pedicle Screw During Posterior Reconstructive SurgerySpine, 2010
- High Strain Rate Testing of Bovine Trabecular BoneJournal of Biomechanical Engineering, 2010
- Biomechanical analysis of differing pedicle screw insertion anglesClinical Biomechanics, 2007
- Instrumentation-Related Complications of Multilevel Fusions for Adult Spinal Deformity Patients Over Age 65Spine, 2006
- A Stronger Bicortical Sacral Pedicle Screw Fixation Through The S1 EndplateSpine, 2005
- The significance of radiolucent zones surrounding pedicle screwsThe Journal of Bone and Joint Surgery. British volume, 2004
- Pedicle Screw Pullout StrengthSpine, 1993
- Importance of Bone Mineral Density in Instrumented Spine FusionsSpine, 1991