Positologica: A Positron ECT Device with a Continuously Rotating Detector Ring
Authors:
- 1 June 1980
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 27 (3) , 1128-1136
Abstract
A rotary positron emission computed tomography device is developed for human brain and animal studies. The device utilizes 64 rectangular BGO detectors arranged at unequal spacing on a circular ring. The detector ring is continuously rotated at a constant speed of 60 rpm or less. This single continuous motion of the ring with the detector array provides excellent sampling characteristics and high detector redundancy. The device has a field of view 24 cm in diameter with a slice thickness adjustable from 1 cm to 2 cm. Measured width of system response to a 2 mm diameter line source is 5.8 mm FWHM at the center and less than 9 mm FWHM within a circle 16 cm in diameter. Measured sensitivity including scattered coincidence events is about 17 kcps/pCi/ml for a 20 cm diameter water phantom and 2 cm thick slice.Keywords
This publication has 19 references indexed in Scilit:
- Data Collection for Cross-Sectional Image Reconstruction by a Moving Ring of Positron Annihilation DetectorsJournal of Computer Assisted Tomography, 1979
- Sampling Requirements and Detector Motion for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1979
- The Mathematics of Wobbling a Ring of Positron Annihilation DetectorsIEEE Transactions on Nuclear Science, 1979
- "Positology"-the search for suitable detector arrangements for a positron ECT with continuous rotationIEEE Transactions on Nuclear Science, 1979
- Engineering Aspects of PETT VIEEE Transactions on Nuclear Science, 1979
- Design and Performance Characteristics of a Positron Emission Computed Axial Tomograph--ECAT®-IIIEEE Transactions on Nuclear Science, 1979
- Design Considerations for a Positron Emission Transverse Tomograph (PETT V) for Imaging of the BrainJournal of Computer Assisted Tomography, 1978
- A Computer Assisted Ringdector Positron Camera System for Reconstruction Tomography of the BrainIEEE Transactions on Nuclear Science, 1978
- Dynamic Positron Emission Tomography for Study of Cerebral Hemodynamics in a Cross Section of the Head Using Positron-Emitting 68Ga-EDTA and 77KrJournal of Computer Assisted Tomography, 1977
- Bismuth Germanate: A High-Z Gamma-Ray and Charged Particle DetectorIEEE Transactions on Nuclear Science, 1975