The design features of cemented femoral hip implants
Authors:
- 1 November 2006
- journal article
- review article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 88-B (11) , 1409-1418
Abstract
We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from in vitro studies and finite element modelling were examined with reference to the survivorship results for each of the concepts of fixation.Keywords
This publication has 128 references indexed in Scilit:
- CT analysis of defects of the cement mantle and alignment of the stemThe Journal of Bone and Joint Surgery. British volume, 2006
- Influence of stem geometry on the stability of polished tapered cemented femoral stemsThe Journal of Bone and Joint Surgery. British volume, 2005
- Spannungsanalyse einer anatomisch adaptierten Femurschaftprothese (Lubinus SPII)Zeitschrift für Orthopädie und ihre Grenzgebiete, 1999
- Failure of Total Hip Arthroplasty With a Precoated ProsthesisClinical Orthopaedics and Related Research, 1998
- Femoral stem fixationThe Journal of Bone and Joint Surgery. British volume, 1998
- Loosening of matt and polished cemented femoral stemsThe Journal of Bone and Joint Surgery. British volume, 1998
- Radiostereometry of Hip ProsthesesClinical Orthopaedics and Related Research, 1997
- In vitro measurement of strain in the bone cement surrounding the femoral component of total hip replacements during simulated gait and stair‐climbingJournal of Orthopaedic Research, 1996
- Survival of the Lubinus Hip ProsthesisClinical Orthopaedics and Related Research, 1994
- "Modes of Failure" of Cemented Stem-type Femoral ComponentsClinical Orthopaedics and Related Research, 1979