A Bland–Altman Comparison of the Lead Care® System and Inductively Coupled Plasma Mass Spectrometry for Detecting Low-Level Lead in Child Whole Blood Samples
Authors:
Open Access
- 1 October 2010
- journal article
- research article
- Published by Springer Nature in Journal of Medical Toxicology
- Vol. 7 (1) , 24-32
Abstract
Chronic childhood lead exposure, yielding blood lead levels consistently below 10 μg/dL, remains a major public health concern. Low neurotoxic effect thresholds have not yet been established. Progress requires accurate, efficient, and cost-effective methods for testing large numbers of children. The LeadCare® System (LCS) may provide one ready option. The comparability of this system to the “gold standard” method of inductively coupled plasma mass spectrometry (ICP-MS) for the purpose of detecting blood lead levels below 10 μg/dL has not yet been examined. Paired blood samples from 177 children ages 5.2–12.8 years were tested with LCS and ICP-MS. Triplicate repeat tests confirmed that LCS and ICP-MS had comparable repeatability. As compared with ICP-MS, LCS had a negative bias of 0.457 μg/dL with an average variability of 1.0 μg/dL. The reproducibility and precision of the LCS is appropriate for the evaluation and monitoring of blood lead levels of individual children in a clinical setting. Recent research however has suggested that increments as small as 0.5 μg/dL may distinguish those at risk of low-level lead-induced neurotoxicity. Thus, we also conclude that the LCS is not useful for research applications attempting to identify neurotoxic effect thresholds for chronic lowest level lead exposure in children. For these types of research applications, a convenient and low-cost device is needed for the precise detection of child blood lead levels below 10 μg/dL.Keywords
This publication has 14 references indexed in Scilit:
- Neuropsychological function in children with blood lead levels <10μg/dLNeuroToxicology, 2007
- Using Bland−Altman to Assess Agreement between Two Medical Devices – Don’t Forget the Confidence Intervals!Journal of Clinical Monitoring and Computing, 2007
- Blood lead levels and specific attention effects in young childrenNeurotoxicology and Teratology, 2007
- Neurobehavioral function in children with low blood lead concentrationsNeuroToxicology, 2007
- A rationale for lowering the blood lead action level from 10 to 2μg/dLNeuroToxicology, 2006
- Blood Lead Concentrations in Maternal and Cord Blood Evaluated by Two Analytic MethodsArchives of Environmental & Occupational Health, 2005
- Neurodevelopmental effects of postnatal lead exposure at very low levelsNeurotoxicology and Teratology, 2004
- Determination of lead in whole blood: Comparison of the LeadCare blood lead testing system with Zeeman longitudinal electrothermal atomic absorption spectrometryJournal of Trace Elements in Medicine and Biology, 2002
- Screen-printed stripping voltammetric/potentiometric electrodes for decentralized testing of trace leadAnalytical Chemistry, 1992
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986