Comparison of Four Different Pediatric 10F Aortic Cannulae During Pulsatile Versus Nonpulsatile Perfusion in a Simulated Neonatal Model of Cardiopulmonary Bypass
Authors:
- Bingyang JiBJBingyang Ji
- Alan RiderARAlan Rider
- Branka LukicBLBranka Lukic
- Allen R. KunselmanAKAllen R. Kunselman
- William J. WeissWWWilliam J. Weiss
- John L. MyersJMJohn L. Myers
- 1 November 2007
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 53 (6) , 778-784
Abstract
We compared four commercially available 10F pediatric aortic cannulae with different geometric designs (DLP-Long tip, DLP-Short tip, RMI-Long tip, and Surgimedics-Short tip) during pulsatile versus nonpulsatile perfusion in terms of pressure drops and surplus hemodynamic energy (SHE) levels in an in vitro neonatal model of cardiopulmonary bypass. The pseudo patient was subjected to seven pump flow rates at 100 ml/min increments in the 400-1,000 ml/min range. A total of 44 experiments (n = 22, nonpulsatile; n = 22, pulsatile) were performed at each of the seven flow rates. Surgimedics had significantly higher pressure drops than the other three cannulae at various flow rates during nonpulsatile and pulsatile perfusion, respectively. When the perfusion mode was changed from nonpulsatile to pulsatile flow, SHE levels at both precannula and postcannula sites increased seven to nine times at all flow rates in all four cannulae. Surgimedics generated a significant lower SHE level when compared with the other three cannulae at all flow rates at both precannula and postcannula sites. The results suggest that different geometries of aortic cannulae have a significant impact on pressure drops of the cannulae as well as hemodynamic energy generation and delivery. Pulsatile perfusion generates more "extra" hemodynamic energy when compared with the nonpulsatile perfusion mode with all four cannulae used in this study.Keywords
This publication has 24 references indexed in Scilit:
- Quantification of Perfusion Modes in Terms of Surplus Hemodynamic Energy Levels in a Simulated Pediatric CPB ModelASAIO Journal, 2006
- An Evaluation of the Benefits of Pulsatile versus Nonpulsatile Perfusion during Cardiopulmonary Bypass Procedures in Pediatric and Adult Cardiac PatientsASAIO Journal, 2006
- Outcomes of the First International Conference on Pediatric Mechanical Circulatory Support Systems and Pediatric Cardiopulmonary PerfusionASAIO Journal, 2006
- Effects of Pulsatile CPB on Interleukin‐8 and Endothelin‐1 LevelsArtificial Organs, 2005
- Pulsatile Versus Nonpulsatile Cardiopulmonary Bypass Procedures in Neonates and Infants: From Bench to Clinical PracticeASAIO Journal, 2005
- Precise Quantification of Pulsatility is a Necessity for Direct Comparisons of Six Different Pediatric Heart-Lung Machines in a Neonatal CPB ModelASAIO Journal, 2005
- Pulsatile Perfusion Improves Regional Myocardial Blood Flow during and after Hypothermic Cardiopulmonary Bypass in a Neonatal Piglet ModelASAIO Journal, 2002
- Monitoring Regional Cerebral Oxygen Saturation Using Near-Infrared Spectroscopy During Pulsatile Hypothermic Cardiopulmonary Bypass in a Neonatal Piglet ModelASAIO Journal, 2000
- Defining Pulsatile Perfusion: Quantification in Terms of Energy Equivalent PressureArtificial Organs, 1999
- Flow Characteristics of Aortic CannulaeJournal of Cardiac Surgery, 1995