Weight-Based, Low-Dose Pediatric Whole-Body PET/CT Protocols
Authors:
- Adam M. AlessioAAAdam M. Alessio
- Paul E. KinahanPKPaul E. Kinahan
- Vivek ManchandaVMVivek Manchanda
- Victor GhioniVGVictor Ghioni
- Lisa AldapeLALisa Aldape
- Marguerite T. ParisiMPMarguerite T. Parisi
Open Access
- 29 September 2009
- journal article
- Published by Society of Nuclear Medicine in Journal of Nuclear Medicine
- Vol. 50 (10) , 1570-1578
Abstract
Adult PET/CT acquisition protocols need to be modified for pediatric imaging to minimize the radiation dose while maintaining diagnostic utility. We developed pediatric PET/CT acquisition protocols customized to patient weight and estimated the dosimetry and cancer risk of these low-dose protocols to communicate basic imaging risks. Methods: Protocols were developed for whole-body 18F-FDG imaging of patients in PET mode with a weight-based injected activity (5.3 MBq/kg) and acquisition times (3–5 min/field of view) and for CT for attenuation correction and localization with a weight-based tube current ranging from 10 to 40 mAs. Patients were categorized on the basis of the Broselow–Luten color-coded weight scale. Dosimetry and radiation-induced cancer risk for the PET and CT acquisition in each category were derived from mean patient sizes and the interpolation of factors from accepted patient models. Results: Whole-body pediatric PET/CT protocols require the customization of PET-acquisition settings and task-specific selection of CT technique. The proposed weight-based protocols result in an approximate effective dose ranging from 8.0 mSv for a 9-kg patient up to 13.5 mSv for a 63-kg patient. The radiation dose from the proposed protocols is 20%−50% (depending on patient weight), the dose from PET/CT protocols that use a fixed CT technique of 120 mAs and 120 kVp. The approximate, conservative estimate of additional lifetime attributable risk (LAR) of cancer incidence for females using the proposed protocols was approximately 3 in 1,000, with a variation of 18% across patient categories. For males, the additional LAR of cancer incidence was approximately 2 in 1,000, with a variation of 16% across categories. Conclusion: Low-dose PET/CT protocols for 11 patient weight categories were developed. The proposed protocols offer an initial set of acquisition parameters for pediatric PET/CT. The use of multiple categories allows for the continued refinement of dose-reduction parameters to minimize dose while maintaining image quality across the range of pediatric patient sizes.Keywords
This publication has 33 references indexed in Scilit:
- Guidelines for 18F-FDG PET and PET-CT imaging in paediatric oncologyEuropean Journal of Nuclear Medicine and Molecular Imaging, 2008
- Guidelines for the use of PET–CT in childrenNuclear Medicine Communications, 2008
- The ‘Image Gently’ campaign: increasing CT radiation dose awareness through a national education and awareness programPediatric Radiology, 2008
- Performance of the new generation of whole-body PET/CT scanners: Discovery STE and Discovery VCTEuropean Journal of Nuclear Medicine and Molecular Imaging, 2007
- PET imaging in pediatric oncologyPediatric Radiology, 2004
- Issues specific to implementing PET?CT for pediatric oncology: what we have learned along the wayPediatric Radiology, 2004
- Pediatric CT: practical approach to diminish the radiation dosePediatric Radiology, 2002
- Improved Pediatric Multidetector Body CT Using a Size-Based Color-Coded FormatAmerican Journal of Roentgenology, 2002
- Risk of Ionizing Radiation Exposure to Children: A Subject ReviewPediatrics, 1998
- Length-based endotracheal tube and emergency equipment in pediatricsAnnals of Emergency Medicine, 1992