Implementation of Sphinx/Lynx as daily QA equipment for scanned proton and carbon ion beams
Open Access
- 27 January 2023
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 24 (4)
Abstract
Purpose: Reporting on the first implementation of a proton dedicated commercial device (IBA Sphinx/Lynx) for daily Quality Assurance (QA) of scanned proton and carbon ion beams.Methods: Daily QA trendlines over more than 3 years for protons and more than 2 years for carbon ions have been acquired. Key daily QA parameters were reviewed, namely the spot size and position, beam range, Bragg peak width, coincidence (between beam and imaging system isocenters), homogeneity and dose.Results: The performance of the QA equipment for protons and carbon ions was evaluated. Daily QA trendlines allowed us to detect machine performance drifts and changes. The definition of tolerances and action levels is provided and compared with levels used in the literature.Conclusion: The device has been successfully implemented for routine daily QA activities in a dual particle therapy facility for more than 2 years. It improved the efficiency of daily QA and provides a comprehensive QA process.Keywords
This publication has 16 references indexed in Scilit:
- Implementation of dosimetry equipment and phantoms at the MedAustron light ion beam therapy facilityMedical Physics, 2017
- A comprehensive and efficient daily quality assurance for PBS proton therapyPhysics in Medicine & Biology, 2017
- Characterization of a commercial scintillation detector for 2-D dosimetry in scanned proton and carbon ion beamsPhysica Medica, 2017
- A fast and reliable method for daily quality assurance in spot scanning proton therapy with a compact and inexpensive phantomMedical Dosimetry, 2017
- Evaluation of beam delivery and ripple filter design for non-isocentric proton and carbon ion therapyPhysics in Medicine & Biology, 2015
- On the nuclear halo of a proton pencil beam stopping in waterPhysics in Medicine & Biology, 2015
- The CNAO dose delivery system for modulated scanning ion beam radiotherapyMedical Physics, 2015
- Dosimetric consequences of pencil beam width variations in scanned beam particle therapyPhysics in Medicine & Biology, 2013
- Dosimetry for ion beam radiotherapyPhysics in Medicine & Biology, 2010
- Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40Medical Physics, 1994