Bi-harmonic cantilever design for improved measurement sensitivity in tapping-mode atomic force microscopy
- 1 April 2014
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 85 (4) , 043703
Abstract
This paper presents a method and cantilever design for improving the mechanical measurement sensitivity in the atomic force microscopy (AFM) tapping mode. The method uses two harmonics in the drive signal to generate a bi-harmonic tapping trajectory. Mathematical analysis demonstrates that the wide-valley bi-harmonic tapping trajectory is as much as 70% more sensitive to changes in the sample topography than the standard single-harmonic trajectory typically used. Although standard AFM cantilevers can be driven in the bi-harmonic tapping trajectory, they require large forcing at the second harmonic. A design is presented for a bi-harmonic cantilever that has a second resonant mode at twice its first resonant mode, thereby capable of generating bi-harmonic trajectories with small forcing signals. Bi-harmonic cantilevers are fabricated by milling a small cantilever on the interior of a standard cantilever probe using a focused ion beam. Bi-harmonic drive signals are derived for standard cantilevers and bi-harmonic cantilevers. Experimental results demonstrate better than 30% improvement in measurement sensitivity using the bi-harmonic cantilever. Images obtained through bi-harmonic tapping exhibit improved sharpness and surface tracking, especially at high scan speeds and low force fields.Keywords
Funding Information
- NSF (CMMI-1229701)
This publication has 28 references indexed in Scilit:
- High-speed atomic force microscopy coming of ageNanotechnology, 2012
- Harnessing the transient signals in atomic force microscopyInternational Journal of Robust and Nonlinear Control, 2005
- A mechanical microscope: High-speed atomic force microscopyApplied Physics Letters, 2005
- Dynamics of the cantilever in noncontact dynamic force microscopy: The steady-state approximation and beyondPhysical Review B, 2001
- High performance feedback for fast scanning atomic force microscopesReview of Scientific Instruments, 2001
- Dissipation processes of an oscillating microlever in non-contact resonant mode: a mechanical approachJournal of Physics D: Applied Physics, 2001
- Atomic resolution on Si(111)-(7×7) by noncontact atomic force microscopy with a force sensor based on a quartz tuning forkApplied Physics Letters, 2000
- Calibration of optical lever sensitivity for atomic force microscopyReview of Scientific Instruments, 1995
- An atomic-resolution atomic-force microscope implemented using an optical leverJournal of Applied Physics, 1989
- Atomic Force MicroscopePhysical Review Letters, 1986