The Impact of Operator Access on Implant Surface Roughness Following Implantoplasty Procedures: A Laboratory Study
Open Access
- 18 March 2026
- journal article
- Published by Wiley in Clinical and Experimental Dental Research
- Vol. 12 (2) , e70336
Abstract
Objectives: To evaluate the influence of operator access on surface roughness parameters and material loss following implantoplasty (IP). Material and Methods: A total of 42 dental implants were utilized in this study, and three different implant designs. Two set‐ups were used: a table‐top, allowing unrestricted access, and a phantom‐head, imposing limitations in access and better replicating the clinical situation. A combination of tungsten carbide burs and silicone polishers was used to perform IP on the implants. The total amount of time required for IP of each implant was measured for comparisons between modalities, and a 2D roughness profile analysis ( R a and R z ) was conducted using a contact stylus profilometer. The weight and the diameter of the implants were also measured before and after IP was performed. Results: A statistically significant difference was found between the R a ( p = 0.004) and R z ( p = 0.019) values of implants prepared under table‐top conditions, being smoother, when compared to the ones prepared in the phantom head. The time required to perform IP was significantly different between the two different settings ( p < 0.001) and between some implant types. The set‐up in which IP was performed did not have a significant impact on the change in implant mass ( p = 0.926) or in implant diameter ( p = 0.721). Conclusions: Implants treated with IP in the phantom‐head set‐up exhibited significantly greater surface roughness and required longer procedural time compared to those treated in the table‐top model, but did not influence material loss.Keywords
This publication has 29 references indexed in Scilit:
- Reduced Biofilm Accumulation on Implants Treated With Implantoplasty—An In Situ Trial With a Within‐Subject ComparisonClinical and Experimental Dental Research, 2024
- Titanium implant surface roughness after different implantoplasty protocols: A laboratory studyClinical and Experimental Dental Research, 2022
- Implantoplasty- provoking or reducing inflammation? – a systematic scoping reviewActa Odontologica Scandinavica, 2021
- Reduced fracture load of dental implants after implantoplasty with different instrumentation sequences. An in vitro studyClinical Oral Implants Research, 2021
- Implantoplasty: Carbide burs vs diamond sonic tips. An in vitro studyClinical Oral Implants Research, 2020
- Does implantoplasty affect the failure strength of narrow and regular diameter implants? A laboratory studyClinical Oral Investigations, 2020
- Mechanical and biological complications after implantoplasty—A systematic reviewClinical Oral Implants Research, 2019
- Comparison of two different techniques used for the maintenance of peri-implant soft tissue health: a pilot randomized clinical trialActa Odontologica Scandinavica, 2017
- Prevalences of peri-implantitis and peri-implant mucositis: systematic review and meta-analysisJournal of Dentistry, 2017
- Surface roughness of dental implants and treatment time using six different implantoplasty proceduresClinical Oral Implants Research, 2015