Temperature change rate actuated bubble mixing for homogeneous rehydration of dry pre-stored reagents in centrifugal microfluidics
- 20 December 2017
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 18 (2) , 362-370
Abstract
In centrifugal microfluidics, dead volumes in valves downstream of mixing chambers can hardly be avoided. These dead volumes are excluded from mixing processes and hence cause a concentration gradient. Here we present a new bubble mixing concept which avoids such dead volumes. The mixing concept employs heating to create a temperature change rate (TCR) induced overpressure in the air volume downstream of mixing chambers. The main feature is an air vent with a high fluidic resistance, representing a low pass filter with respect to pressure changes. Fast temperature increase causes rapid pressure increase in downstream structures pushing the liquid from downstream channels into the mixing chamber. As air further penetrates into the mixing chamber, bubbles form, ascend due to buoyancy and mix the liquid. Slow temperature/pressure changes equilibrate through the high fluidic resistance air vent enabling sequential heating/cooling cycles to repeat the mixing process. After mixing, a complete transfer of the reaction volume into the downstream fluidic structure is possible by a rapid cooling step triggering TCR actuated valving. The new mixing concept is applied to rehydrate reagents for loop-mediated isothermal amplification (LAMP). After mixing, the reaction mix is aliquoted into several reaction chambers for geometric multiplexing. As a measure for mixing efficiency, the mean coefficient of variation ( , n = 4 LabDisks) of the time to positivity (tp) of the LAMP reactions (n = 11 replicates per LabDisk) is taken. The of the tp is reduced from 18.5% (when using standard shake mode mixing) to 3.3% (when applying TCR actuated bubble mixing). The bubble mixer has been implemented in a monolithic fashion without the need for any additional actuation besides rotation and temperature control, which are needed anyhow for the assay workflow.Keywords
Funding Information
- Seventh Framework Programme (318408)
- Horizon 2020 Framework Programme (688207)
- Deutsche Forschungsgemeinschaft (ZE 527/10-1)
This publication has 33 references indexed in Scilit:
- Fully integrated, fully automated generation of short tandem repeat profilesInvestigative Genetics, 2013
- Vacuum/Compression Valving (VCV) Using Parrafin-Wax on a Centrifugal Microfluidic CD PlatformPLOS ONE, 2013
- Centrifugal gas-phase transition magnetophoresis (GTM) – a generic method for automation of magnetic bead based assays on the centrifugal microfluidic platform and application to DNA purificationLab on a Chip, 2012
- Lab-on-a-Chip Pathogen Sensors for Food SafetySensors, 2012
- Fiji: an open-source platform for biological-image analysisNature Methods, 2012
- Aliquoting on the centrifugal microfluidic platform based on centrifugo-pneumatic valvesMicrofluidics and Nanofluidics, 2011
- Infrared controlled waxes for liquid handling and storage on a CD-microfluidic platformLab on a Chip, 2010
- Characterizing dispersion in microfluidic channelsLab on a Chip, 2009
- Highly accurate sigmoidal fitting of real-time PCR data by introducing a parameter for asymmetryBMC Bioinformatics, 2008
- One-step pathogen specific DNA extraction from whole blood on a centrifugal microfluidic deviceLab on a Chip, 2007