Three-dimensional cardiac rotational angiography: effective radiation dose and image quality implications
Open Access
- 17 December 2009
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 12 (2) , 194-201
Abstract
Three-dimensional rotational angiography (3DRA) is a promising new online tool for 3D imaging during cardiac ablation procedures. No precise data exist concerning its associated radiation dose. The current study evaluated the effective dose (ED) of cardiac rotational angiography and its relation to patient properties, imaging system input settings, and quality of reconstructed 3D images. We performed Monte Carlo simulation-based radiation dose calculations in 42 patients referred for ablation of cardiac arrhythmias. Detailed tube setting information from the 3DRA system (Siemens Axiom Artis dBC with Syngo DynaCT Cardiac software) was used to provide an accurate input for dose calculations in all 248 frames used during image acquisition. Our calculations yielded an overall mean ED of 6.6 ± 1.8 mSv (based on ICRP 103 weighing factors). Manual collimation of the radiation beam can reduce ED by more than 20%. Image quality did not significantly relate to patient body mass index (BMI), dose per frame setting, or dose–area product (DAP), but was rather explained by contrast filling, cardiac motion reduction, and absence of image reconstruction artefacts. In the system evaluated, DAP values are nearly independent from BMI ( R2 = 0.30), due to its technical specifications. Therefore, patient BMI showed an unexpected strong inverse relation to ED. Three-dimensional rotational angiography can be performed with acceptable patient radiation dose, comparable to cardiac CT. With the 3DRA system studied (Siemens Axiom), slender patients may currently receive unnecessarily high radiation doses when compared with obese patients, so that further dose reduction seems feasible for many patients. Adequate collimation is imperative to limit patient exposure.Keywords
This publication has 20 references indexed in Scilit:
- Intraprocedural reconstruction of the left atrium and pulmonary veins as a single navigation tool for ablation of atrial fibrillation: A feasibility, efficacy, and safety studyHeart Rhythm, 2009
- Adenosine-induced ventricular asystole or rapid ventricular pacing to enhance three-dimensional rotational imaging during cardiac ablation proceduresEP Europace, 2009
- Image Integration‐Guided Catheter Ablation of Atrial Fibrillation: A Prospective Randomized StudyJournal of Cardiovascular Electrophysiology, 2009
- Biplane three-dimensional augmented fluoroscopy as single navigation tool for ablation of atrial fibrillation: Accuracy and clinical valueHeart Rhythm, 2008
- Computed tomography-fluoroscopy overlay evaluation during catheter ablation of left atrial arrhythmiaEP Europace, 2008
- Radiation Dose to Patients From Cardiac Diagnostic ImagingCirculation, 2007
- Obesity Is a Major Determinant of Radiation Dose in Patients Undergoing Pulmonary Vein Isolation for Atrial FibrillationJACC, 2007
- Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily PracticeCirculation, 2006
- Validation of Three‐Dimensional Cardiac Image Integration: Use of Integrated CT Image into Electroanatomic Mapping System to Perform Catheter Ablation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2006
- Integrated Electroanatomic Mapping With Three-Dimensional Computed Tomographic Images for Real-Time Guided AblationsCirculation, 2006