Abstract
The error in prediction of emmetropic intraocular lens power or postoperative refractive error after lens implantation was analyzed in three groups of eyes after posterior chamber lens implantation. Regression line calculation with the SRK equation or with a groupspecific regression was compared with theoretical calculations in unselected, long myopic and short hyperopic eyes. The cut-off length was below 22.0 mm for the short eyes and above 25.9 mm for the long eyes. In the unselected and hyperopic group, there was only a small difference in mean error and error variance when the three calculation methods were compared. In the high myopic group, the range of error increased in all methods. The worst results were obtained with the standard SRK equation because the slope of the regression line in myopic eyes differs from the classical regression line calculated on an average population of implants. Lens calculation in high myopic eyes should therefore be performed with a specific regression line or by theoretical calculation.Keywords
This publication has 9 references indexed in Scilit:
- Theoretical, computer-assisted prediction versus SRK prediction of postoperative refraction after intraocular lens implantationJournal of Cataract & Refractive Surgery, 1987
- Analysis of intraocular lens power calculationAmerican Intra-Ocular Implant Society Journal, 1985
- Clinical evaluation of six intraocular lens calculation formulasAmerican Intra-Ocular Implant Society Journal, 1985
- Preoperative evaluation of the cataractous patientSurvey of Ophthalmology, 1984
- The fudged formula for intraocular lens power calculationsAmerican Intra-Ocular Implant Society Journal, 1982
- Axial length measurement and its relation to intraocular lens power calculationsAmerican Intra-Ocular Implant Society Journal, 1982
- Myopic astigmatism a substitute for accommodation in pseudophakiaDocumenta Ophthalmologica, 1981
- A new intraocular lens calculation formulaAmerican Intra-Ocular Implant Society Journal, 1980
- Biometric analysis of intraocular lens power required to produce emmetropia: results of 450 implantsAmerican Intra-Ocular Implant Society Journal, 1978