Abstract
Introduction: Recent neurosurgical applications based on artificial intelligence (AI) have demonstrated its potential in surgical planning and anatomical measurement. We aimed to evaluate the performance of an AI planning software application on screw length/diameter selection and insertion accuracy in comparison with freehand surgery.Methods: A total of 45 patients with 208 pedicle screw placements on thoracolumbar segments were included in this analysis. The novel AI planning software was developed based on a deep learning model. AI-based pedicle screw placements were selected on the basis of preoperative computed tomography (CT) data, and freehand surgery screw placements were observed based on postoperative CT data. The performance of AI pedicle screw placements was evaluated on the components of screw length, diameter, and Gertzbein grade in comparison with the results achieved by freehand surgery.Results: Among 208 pedicle screw placements, the average screw length/diameters selected by the AI model and used in freehand surgery were 48.65 ± 5.99 mm/7.39 ± 0.42 mm and 44.78 ± 2.99 mm/6.1 ± 0.27 mm, respectively. Among AI screw placements, 85.1% were classified as Gertzbein Grade A (no cortical pedicle breach); among free-hand surgery placements, 64.9% were classified as Gertzbein Grade A.Conclusion: The novel AI planning software application could provide an accessible and safe pedicle screw placement strategy in comparison with traditional freehand pedicle screw placement strategies. The choices of pedicle screw dimensional parameters made by the model, including length and diameter, may provide potential inspiration for real clinical discretion.Keywords
This publication has 19 references indexed in Scilit:
- A retrospective study to validate an intraoperative robotic classification system for assessing the accuracy of kirschner wire (K-wire) placements with postoperative computed tomography classification system for assessing the accuracy of pedicle screw placementsMedicine, 2016
- Assessing the Intraoperative Accuracy of Pedicle Screw Placement by Using a Bone-Mounted Miniature Robot System through Secondary RegistrationPLOS ONE, 2016
- Pedicle Morphometry of Lumbar VertebraeSpine, 2015
- Techniques and accuracy of thoracolumbar pedicle screw placementWorld Journal of Orthopedics, 2014
- Pedicle screw diameter selection for safe insertion in the thoracic spineEuropean Journal of Orthopaedic Surgery & Traumatology, 2011
- A computed tomography–based morphometric study of thoracic pedicle anatomy in a random United States trauma populationJournal of Neurosurgery: Spine, 2011
- Accuracy of Free-Hand Pedicle Screws in the Thoracic and Lumbar Spine: Analysis of 6816 Consecutive ScrewsNeurosurgery, 2011
- Diameter, Length, and Direction of Pedicle Screws for Scoliotic SpineSpine, 2009
- Thoracic Pedicle Screw Fixation in Spinal DeformitiesSpine, 2001
- Factors Affecting the Pullout Strength of Cancellous Bone ScrewsJournal of Biomechanical Engineering, 1996