Effects of applied contact force and volume control setting on output force levels of the BAHA® Softband
- 1 January 2006
- journal article
- research article
- Published by Taylor & Francis in International Journal of Audiology
- Vol. 45 (5) , 301-308
Abstract
The BAHA® Softband has been developed to provide a transcutaneous anchor for a BAHA® until a child is a surgical candidate for the percutaneous BAHA® implant. We tested the objective output force level of the BAHA® Classic 300 and Compact connected to a Softband on an artificial mastoid to determine: (1) the effects of direct contact force on output force levels (dB); and (2) the required volume control setting to ensure audibility of speech (assuming an average adult reference equivalent threshold force level). Direct contact force was varied from 2 to 5 N in 1 N steps. Output force level increased with increasing contact force. However, the average increase was 3 dB or lower, suggesting that the contact force is of minor importance. Volume control setting appears to be of much greater importance. It is suggested that the volume setting of either device be set to at least 2.5 to ensure audibility of conversational speech. Data from normal-hearing adults with simulated conductive hearing losses are presented to validate this conclusion.Keywords
This publication has 19 references indexed in Scilit:
- Candidacy for the Bone-Anchored Hearing AidAudiology and Neurotology, 2004
- Limitations and Uses of the Aided AudiogramSeminars in Hearing, 2004
- Is functional gain really functional?The Hearing Journal, 2002
- Osseointegrated Implants in Children: Experience from Our First 100 PatientsOtolaryngology–Head and Neck Surgery, 2001
- Bone-anchored hearing aidsOtolaryngologic Clinics of North America, 2001
- Comparison of Severely and Profoundly Hearing-Impaired Children's Amplification Preferences with the NAL-RP and the DSL 3.0 PrescriptionsScandinavian Audiology, 1997
- Prediction of Hearing Aid BenefitEar & Hearing, 1994
- Distribution of short-term rms levels in conversational speechThe Journal of the Acoustical Society of America, 1988
- Effects of Stimulus Intensity on Speech Perception by Deaf ChildrenJournal of Speech and Hearing Disorders, 1977
- Audiologic Evaluation of Deaf ChildrenJournal of Speech and Hearing Disorders, 1976