Orthorhombic Non‐Perovskite CsPbI3 Microwires for Stable High‐Resolution X‐Ray Detectors
- 1 May 2022
- journal article
- research article
- Published by Wiley in Advanced Optical Materials
- Vol. 10 (13)
Abstract
X‐ray detectors are used widely, but they suffer from short operational stability and insufficient absorptivity of X‐ray photons. Here, a strategy for room‐temperature, solution‐grown δ‐CsPbI3 monocrystalline microwires exhibiting one of the highest linear X‐ray absorption coefficients among known semiconductors is reported. In a metal–semiconductor–metal architecture, δ‐CsPbI3 demonstrates one of the lowest detectable X‐ray dose rates at 33.3 nGyair s−1, enabled by its exceptionally low dark current density of 12 pA mm−2. The detector remains stable following a continuous operation for a total accumulated dose of 1.44 × 106 Gyair, which is equivalent to 720 million chest X‐ray scans. The 1D crystalline nature of δ‐CsPbI3 enables a high X‐ray imaging resolution of 12.4 line pairs per millimeter (lp mm−1).Keywords
This publication has 46 references indexed in Scilit:
- Use of active personal dosemeters in interventional radiology and cardiology: Tests in laboratory conditions and recommendations - ORAMED projectRadiation Measurements, 2011
- Anomalous segregation during electrodynamic gradient freeze growth of cadmium zinc tellurideJournal of Crystal Growth, 2011
- Amorphous and Polycrystalline Photoconductors for Direct Conversion Flat Panel X-Ray Image SensorsSensors, 2011
- X-ray sensitivity of Cd0.9Zn0.1Te detectorsTechnical Physics, 2010
- High-temperature structural evolution of caesium and rubidium triiodoplumbatesJournal of Physics and Chemistry of Solids, 2008
- Cadmium zinc telluride and its use as a nuclear radiation detector materialMaterials Science and Engineering: R: Reports, 2001
- X-ray sensitivity of photoconductors: application to stabilized a-SeJournal of Physics D: Applied Physics, 2000
- Dose Rate Limitations of Integrating Survey Meters for Diagnostic X-Ray SurvevsHealth Physics, 2000
- X-ray detectors for digital radiographyPhysics in Medicine & Biology, 1997
- Digital skeletal radiography: spatial resolution requirements for detection of subperiosteal resorptionAmerican Journal of Roentgenology, 1989