Efficiency of whole-body counter for various body size calculated by MCNP5 software
- 23 August 2012
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 152 (1-3) , 179-183
Abstract
The efficiency of a whole-body counter for 137Cs and 40K was calculated using the MCNP5 code. The ORNL phantoms of a human body of different body sizes were applied in a sitting position in front of a detector. The aim was to investigate the dependence of efficiency on the body size (age) and the detector position with respect to the body and to estimate the accuracy of real measurements. The calculation work presented here is related to the NaI detector, which is available in the Serbian Whole-body Counter facility in Vinca Institute.Keywords
This publication has 9 references indexed in Scilit:
- Application of MCNP5 Software for Efficiency Calculation of a Whole Body CounterHealth Physics, 2012
- Monte Carlo calculations for efficiency calibration of a whole-body monitor using BOMAB phantoms of different sizesRadiation Protection Dosimetry, 2011
- Monte Carlo simulation of the movement and detection efficiency of a whole-body counting system using a BOMAB phantomRadiation Protection Dosimetry, 2011
- Comparison of various anthropomorphic phantom types for in vivo measurements by means of Monte Carlo simulationsRadiation Protection Dosimetry, 2010
- Input files with ORNL—mathematical phantoms of the human body for MCNP-4BComputer Physics Communications, 2007
- Application of voxel phantoms and Monte Carlo method to whole-body counter calibrationRadiation Protection Dosimetry, 2006
- EXTERNAL DOSES TO HUMANS FROM 137Cs IN SOILHealth Physics, 2006
- Individual monitoring for internal exposure in Europe and the integration of dosimetric dataRadiation Protection Dosimetry, 2004
- A catalogue of dosemeters and dosimetric services within Europe--an updateRadiation Protection Dosimetry, 2004