A Study on Characterization of Impulses Pressure Generator by a Drop Weight Method
Open Access
- 1 August 2018
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 1065 (16) , 162012
Abstract
Some experimental approaches were conducted to analyse the characteristics of impulse pressure generated by a drop-weight system, which could be used for the establishment of the dynamic pressure standard. A few parameters such as a drop-height (speed), a drop-weight (force) and a diameter of a Piston-Cylinder Assembly (PCA) have been considered to distinguish these impulse pressures. The impulse pressure system with 8 mm diameter of a PCA could generate fast impulse pressure with duration of less than 3 ms. Similarly, the higher drop-height, the greater impulse pressure. The larger drop-weight increases the pressure peaks, but also, it increases the duration time, too. Since the shortest impulse pressure is preferred in the dynamic pressure calibration system, therefore the larger diameter PCA should be utilized with a constant drop-weight and the peak pressure should be controlled with variable drop-heights. In addition, this paper also describes a simple method to reduce unwanted ripple over the rising impulse that may be occurred in the system. A further simulation is still being conducted to evaluate the frequency source.Keywords
This publication has 3 references indexed in Scilit:
- Dynamic calibration of piezoelectric transducers for ballistic high-pressure measurementInternational Journal of Metrology and Quality Engineering, 2016
- High dynamic pressure standard based on the density change of the step pressure generatorMetrologia, 2013
- Traceable dynamic measurement of mechanical quantities: objectives and first results of this european projectInternational Journal of Metrology and Quality Engineering, 2012