REASSESSMENT OF LUNG-VOLUME MEASUREMENT BY HELIUM DILUTION AND BY BODY PLETHYSMOGRAPHY IN CHRONIC AIR-FLOW OBSTRUCTION
Authors:
- DO RODENSTEINDRDO RODENSTEIN
- DC STANESCUDSDC STANESCU
- 1 January 1982
- journal article
- research article
- Published by Elsevier
- Vol. 126 (6) , 1040-1044
Abstract
In some asthmatic patients, plethysmographic total lung capacity, measured from mouth pressure vs. plethysmographic volume (Vbox) changes (TLCm), is overestimated; TLCes, derived from esophageal pressure vs. Vbox changes, is an accurate estimate of lung volume (VL). Patients (24) with chronic airflow obstruction (mean forced expiratory volume in 1 s, 1.23 l (SD: 0.64 l) and mean specific airway conductance, SGaw, 0.052 .+-. 0.016 cm H2O-1 .cntdot. s-1) were studied to reassess the relative merits of TLCm and the 7-min closed-circuit He dilution technique (TLCHe) in this condition. The TLCHe (6.54 .+-. 1.09 l) was significanlty less (P < 0.001) than both TLCm (7.53 .+-. 1.20 l) and TLCes (7.16 .+-. 1.12 l); TLCm was greater than TLCes (P < 0.001). Differences between TLCes and TLCHe (ranging form -0.11-2.17 l) and between TLCm and TLCes (from -0.10-1.48 l) were correlated to the degree of airway obstruction as assessed by SGaw (P < 0.001 and P < 0.01, respectively). VL is overestimated by TLCm and underestimated by TLCHe in moderate to severe chronic airlow obstruction. The trapped gas volume (TLCm-TLCHe difference) is made-up by these 2 additive errors.Keywords