Assessment of Becton Dickinson Plain and Serum Separator Tubes in Measurement of 25‐Hydroxyvitamin D3 (25OHD3) by HPLC and Immunoassay Methods
Authors:
- 2 October 2014
- journal article
- research article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- Vol. 30 (1) , 32-35
Abstract
Background: The accuracy of 25‐hydroxyvitamin D3 (25OHD3) measurement on specimens collected into serum separator tubes (SSTs) has been questioned because of possible interference by the gel. Possible interference was investigated in SSTs from Becton Dickinson (BD).Design and methods: Blood specimens were collected simultaneously from 50 normal subjects into plain tubes and SSTs. 25OHD3 was assayed on serum using high performance liquid chromatography (Chromsystems), and Architect (Abbott) and Liaison (Diasorin) immunoassays.Results: There were no significant differences between 25OHD3 results (means ± SE, nmol/l) obtained from specimens collected into plain tubes and SSTs assayed by HPLC (39.0 ± 2.7 vs. 39.3 ± 2.7), Liaison (32.9 ± 2.2 vs. 32.8 ± 2.3), or Architect (43.1 ± 2.8 vs. 43.2 ± 2.8). In specimens collected into plain tubes and SSTs, 25OHD3 measurements by HPLC correlated significantly (P < 0.0001) with those from the Architect (r = 0.895, r = 0.908) and Liaison (r = 0.907, r = 0.913), respectively.Conclusions: The gel in SSTs (BD) does not interfere with the measurement of 25OHD3 by HPLC or common immunoassays. This important finding may enable clinical laboratories to make cost savings by using SSTs without concerns about inaccuracy.Keywords
Funding Information
- King Abdullah International Medical Research Center
This publication has 17 references indexed in Scilit:
- State-of-the-Art Vitamin D Assays: A Comparison of Automated Immunoassays with Liquid Chromatography–Tandem Mass Spectrometry MethodsClinical Chemistry, 2012
- Accuracy of 6 Routine 25-Hydroxyvitamin D Assays: Influence of Vitamin D Binding Protein ConcentrationClinical Chemistry, 2012
- Pathology Consultation on Vitamin D TestingAmerican Journal of Clinical Pathology, 2011
- Vitamin D in health and disease: Current perspectivesNutrition Journal, 2010
- Measurement of 25-hydroxyvitamin D in the clinical laboratory: Current procedures, performance characteristics and limitationsSteroids, 2010
- Preanalytical Stability of 25(OH)–Vitamin D3 in Human Blood or Serum at Room Temperature: Solid as a RockClinical Chemistry, 2009
- 25- Hydroxyvitamin D: Explosion in Clinical Interest and Laboratory RequestsOman Medical Journal, 2009
- Vitamin D DeficiencyThe New England journal of medicine, 2007
- HPLC Method for 25-Hydroxyvitamin D Measurement: Comparison with Contemporary AssaysClinical Chemistry, 2006
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986