Diffusion‐weighted MRI of the lung with hyperpolarized helium‐3: A study of reproducibility

Authors:
  • JSJörg Schmiedeskamp
    Jörg Schmiedeskamp

Abstract
Purpose: To determine the reproducibility of several parameters of the ADC measurement by calculating the scan‐to‐scan intrasubject variability.Materials and Methods: Measurements were performed using a gradient‐echo sequence with a bipolar gradient for diffusion weighting (b = 3.89 sec/cm2). Five patients with pulmonary emphysema, and six healthy‐lung volunteers were included in the study. Images were acquired after inspiration of3He during a single inspiratory breath‐hold. To assess the reproducibility, the measurement was performed twice (time between measurements = 20 minutes) without repositioning the subjects. Analysis was performed on the basis of region‐of‐interest (ROI) analysis and global lung ADC histograms.Results: The mean ADC of a ROI varied by 5.1% between two measurements for volunteers and by 6.1% for patients. In the global evaluation, the 75th percentile demonstrated the best reproducibility (2%), while other parameters showed variations up to 12%. Only the variation of the standard deviation (SD) and the measure of homogeneity of the ADC map showed a significant difference between patients and volunteers.Conclusion: Diffusion‐weighted imaging (DWI) is a well‐reproducible method for assessing the lung microstructure. J. Magn. Reson. Imaging 2005;21:765–774.