Intrarater reliability of the Humac NORM isokinetic dynamometer for strength measurements of the knee and shoulder muscles
Open Access
- 10 January 2018
- journal article
- research article
- Published by Springer Nature in BMC Research Notes
- Vol. 11 (1) , 15
Abstract
To determine the intrarater reliability of the Humac NORM isokinetic dynamometer for concentric and eccentric strength tests of knee and shoulder muscles. 54 participants (50% female, average age 20.9 ± 3.1 years) performed concentric and eccentric strength measures of the knee extensors and flexors, and the shoulder internal and external rotators on two different Humac NORM isokinetic dynamometers, which were situated at two different centers. The knee extensors and flexors were tested concentrically at 60° and 180°/s, and eccentrically at 60° s. Concentric strength of the shoulder internal and external rotators, and eccentric strength of the external rotators were measured at 60° and 120°/s. We calculated intraclass correlation coefficients (ICCs), standard error of measurement, standard error of measurement expressed as a %, and the smallest detectable change to determine reliability and measurement error. ICCs for the knee tests ranged from 0.74 to 0.89, whereas ICC values for the shoulder tests ranged from 0.72 to 0.94. Measurement error was highest for the concentric test of the knee extensors and lowest for the concentric test of shoulder external rotators.Keywords
This publication has 21 references indexed in Scilit:
- Rating the methodological quality in systematic reviews of studies on measurement properties: a scoring system for the COSMIN checklistQuality of Life Research, 2011
- Hand‐held Dynamometry Correlation With the Gold Standard Isokinetic Dynamometry: A Systematic ReviewPM&R, 2011
- Isokinetic assessment of the shoulder rotators: a study of optimal test positionClinical Physiology and Functional Imaging, 2011
- Technical Report: Reliability and Validity of the iSAM 9000 Isokinetic DynamometerJournal of Strength & Conditioning Research, 2008
- Reliability of isokinetic strength imbalance ratios measured using the Cybex NORM dynamometerClinical Physiology and Functional Imaging, 2007
- Reliability of knee extension and flexion measurements using the Con‐Trex isokinetic dynamometerClinical Physiology and Functional Imaging, 2007
- When to use agreement versus reliability measuresJournal of Clinical Epidemiology, 2006
- Test–retest reliability of glenohumeral internal and external rotation strength in chronic rotator cuff pathologyPhysical Therapy in Sport, 2006
- Learning effect of isokinetic measurements in healthy subjects, and reliability and comparability of Biodex and Lido dynamometersClinical Physiology and Functional Imaging, 2005
- Isokinetic Eccentric ExerciseSports Medicine, 1995