Hemolysis in a continuous-flow ventricular assist device with/without chamfer
Authors:
Open Access
- 18 April 2017
- journal article
- research article
- Published by SAGE Publications in Advances in Mechanical Engineering
- Vol. 9 (4)
Abstract
Ventricular assist device has been increasingly used in the treatment of end-stage heart failure, with a high incidence of hemolysis in clinical use, which is a serious problem to be solved. In this study, a complete set of continuous-flow ventricular assist device in vitro test platform for hydraulic performance and hemolysis was established. A self-developed continuous-flow ventricular assist device was employed, with the impeller being chamfered or unchamfered, respectively, to undergo the hydraulic performance and in vitro hemolysis tests, using computational fluid dynamics and a rheometer to investigate the influences of the shear stress on hemolysis. Computational fluid dynamics results showed that the shear stress of continuous-flow ventricular assist device with chamfered impeller was significantly decreased; as a result, the normalized index of hemolysis of the chamfered was reduced, demonstrating that lowering the shear stress of the impeller could significantly reduce hemolysis. Moreover, it was further confirmed by a fixed shear stress test. When the shear stress was greater than 100 Pa, the hemolytic effects were increased significantly. In standard operating conditions, lowering the impeller’s shear stress to less than 100 Pa could effectively reduce the hemolytic effects of ventricular assist device.Keywords
Funding Information
- The independent research fund of Tsinghua University (No. 20141081265)
This publication has 27 references indexed in Scilit:
- A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis IndexJournal of Biomechanical Engineering, 2012
- Second INTERMACS annual report: More than 1,000 primary left ventricular assist device implantsThe Journal of Heart and Lung Transplantation, 2010
- Incidence and Patterns of Adverse Event Onset During the First 60 Days After Ventricular Assist Device ImplantationThe Annals of Thoracic Surgery, 2009
- Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist DeviceJACC, 2009
- HeartMate II left ventricular assist device; early European experienceEuropean Journal of Cardio-Thoracic Surgery, 2008
- New Centrifugal Blood Pump With Dual Impeller and Double Pivot Bearing System: Wear Evaluation in Bearing System, Performance Tests, and Preliminary Hemolysis TestsArtificial Organs, 2008
- EXPERIMENTAL FLUID MECHANICS OF PULSATILE ARTIFICIAL BLOOD PUMPSAnnual Review of Fluid Mechanics, 2006
- Cardiac Replacement with a Total Artificial Heart as a Bridge to TransplantationNew England Journal of Medicine, 2004
- Species Differences in Erythrocyte Mechanical FragilityASAIO Journal, 1998
- Experimental determination of the kinematic viscosity of glycerol-water mixturesProceedings of the Royal Society of London. Series A: Mathematical and Physical Sciences, 1994