Technical Feasibility and Safety of a Novel Device for Simultaneous Double Guidewire Insertion and Tract Dilation During Interventional EUS
Authors:
- 25 July 2025
- journal article
- Published by Springer Nature in Digestive Diseases and Sciences
- Vol. 70 (12) , 4229-4234
Abstract
Interventional endoscopic ultrasound (I-EUS) has been developed as an alternative technique. To prevent these adverse events, sufficient tract dilation using by one device or double guidewire technique (DGWT) is useful. For application of the DGWT, we developed a novel double lumen dilator (DLD). This study evaluated the technical feasibility of I-EUS using the DLD. The primary outcome of this study was the technical success rate of DLD insertion into the target site. The secondary outcomes were adverse events associated with the procedure. A total of consecutive 19 patients who underwent I-EUS using DLD as dilation device between November 2024 and December 2024 were enrolled in this study. The technical success rate of DLD insertion was 94.7% (18/19). Mean procedure time was 14.4 min. Among the 18 patients who underwent successful DLD insertion, stent deployment was successfully performed. Stent misdeployment was observed in one patient during EUS-guided hepaticogastrostomy, and the dislocated metal stent could not recovered. However, a plastic stent was successfully deployed by the DGWT. The DLD allows DGWT and tract dilation to be performed as a single step, and therefore, might play a role in the safety and efficacy of the I-EUS procedure.Keywords
This publication has 16 references indexed in Scilit:
- Efficacy and safety of endoscopic ultrasound-guided hepaticogastrostomy: a meta-regression analysisEndoscopy, 2024
- Endoscopic Ultrasound-Guided Pancreatic Pseudocyst Drainage Using a Novel Dilation Device with Double Guidewire Insertion Function (with Video)Digestive Diseases and Sciences, 2024
- Clinical evaluation of a novel drill dilator as the first-line tract dilation technique during EUS-guided biliary drainage by nonexpert hands (with videos)Gastrointestinal Endoscopy, 2023
- Double guidewire technique stabilization procedure for endoscopic ultrasound-guided hepaticogastrostomy involving modifying the guidewire angle at the insertion siteSurgical Endoscopy, 2022
- Double Guidewire Technique Using an Uneven Double Lumen Catheter for Endoscopic Ultrasound-Guided InterventionsDigestive Diseases and Sciences, 2020
- Risk factors for adverse events associated with bile leak during EUS-guided hepaticogastrostomyEndoscopic Ultrasound, 2020
- Double‐guidewire technique for endoscopic ultrasound‐guided pancreatic duct drainageDigestive Endoscopy, 2019
- Safety and efficacy of ultra-tapered mechanical dilator for EUS-guided hepaticogastrostomy and pancreatic duct drainage compared with electrocautery dilator (with video)Endoscopic Ultrasound, 2018
- EUS-guided biliary drainage with transluminal stenting after failed ERCP: predictors of adverse events and long-term resultsGastrointestinal Endoscopy, 2011
- A lexicon for endoscopic adverse events: report of an ASGE workshopGastrointestinal Endoscopy, 2010