Undesirable nuclear reactions and induced radioactivity as a result of the use of the high-energy therapeutic beams generated by medical linacs
Authors:
- 11 June 2007
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 128 (2) , 133-145
Abstract
In this paper, the problem of undesirable photonuclear, electronuclear and neutron capture reactions taking place in the treatment room during emission of the typical high-energy therapeutic beams from two different medical accelerators, i.e. Primus Siemens and Varian Clinac-2300, is presented. The radioisotopes 187W, 56Mn, 28Al, 57Ni, 38Cl, 57Co and 19Au and the neutron activation of 1H were identified as a consequence of these reactions. Moreover, the increased photon fluence rate behind the door of the accelerator bunker in the operator console room was observed during emission of the 20 MV X-rays from the Varian Clinac-2300 as well as in the case of the 15 MV X-ray beam from the Primus Siemens. No increased radiation was observed during the 6 MV X-ray beam emission. The performed measurements produced evidences on the presence of neutrons in the operator console room during emission of the 15 MV X-ray beam from the Primus Siemens as well as the 20 MV X-rays and the 22 MeV electrons from the Varian Clinac-2300 accelerator.Keywords