Laser induced fluorescence in Ar and He plasmas with a tunable diode laser
Authors:
- 1 October 2003
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 74 (10) , 4352-4360
Abstract
A diode laser based laser induced fluorescence (LIF) diagnostic that uses an inexpensive diode laser system is described. This LIF diagnostic has been developed on the hot helicon experiment (HELIX) plasma device. The same diode laser is used to alternatively pump Ar II and He I transitions to obtain argon ion and atomic helium temperatures, respectively. The 1.5 MHz bandwidth diode laser has a Littrow external cavity with a mode-hop free tuning range up to 14 GHz and a total power output of about 12 mW. Wavelength scanning is achieved by varying the voltage on a piezoelectric controlled grating located within the laser cavity. The fluorescence radiation is monitored with a photomultiplier detector. A narrow band interference filter is used to eliminate all but the plasma radiation in the immediate vicinity of the fluorescence wavelength. Lock-in amplification is used to isolate the fluorescence signal from noise and electron-impact induced radiation. For the Ar ion, the laser tuned at 668.43 nm is used to pump the Ar II metastable level to the level. The 442.60 nm fluorescence radiation between the and the levels is captured by the photomultiplier tube. For atomic He, the laser is tuned at 667.82 nm to pump a fraction of the electron population from the state to the upper level. Although the level is not a metastable, the close proximity of metastable makes this new He I LIF scheme possible. In this scheme, a fraction of the laser-excited electrons undergo collisional excitation transfer from the to the level. In turn, the state decays to the metastable by emitting 501.57 nm fluorescence photons.
Keywords
This publication has 30 references indexed in Scilit:
- Ion heating in the HELIX helicon plasma sourcePhysics of Plasmas, 1999
- Control of ion temperature anisotropy in a helicon plasmaPlasma Sources Science and Technology, 1998
- Laser induced fluorescence diagnostic for measuring small magnetic fieldsReview of Scientific Instruments, 1997
- Single beam laser induced fluorescence technique for plasma transport measurementsReview of Scientific Instruments, 1996
- Observation of fast stochastic ion heating by drift wavesPhysical Review Letters, 1987
- Ion Heating Due to Rotation and Collision in Magnetized PlasmaPhysical Review Letters, 1986
- Experimental Measurements of Phase SpacePhysical Review Letters, 1986
- Single frequency scanning laser as a plasma diagnosticReview of Scientific Instruments, 1983
- Direct ion-transport measurement by optical taggingPhysics Letters A, 1983
- Plasma Ion Diagnostics Using Resonant FluorescencePhysical Review Letters, 1975