A continuous size-dependent particle separator using a negative dielectrophoretic virtual pillar array
Authors:
- 4 September 2008
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 8 (11) , 1930-1936
Abstract
We present a continuous size-dependent particle separator using a negative dielectrophoretic (DEP) virtual pillar array. Two major problems in the previous size-dependent particle separators include the particle clogging in the mechanical sieving structures and the fixed range of separable particle sizes. The present particle separator uses the virtual pillar array generated by negative DEP force instead of the mechanical pillar array, thus eliminating the clogging problems. It is also possible to adjust the size of separable particles since the size of virtual pillars is a function of a particle diameter, applied voltage, flow rate, etc. At an applied voltage of 500 kHz, 10 Vrms (root mean sqaure voltage) sinusoidal wave and a flow rate of 0.40 μl min−1, we separate 5.7 ± 0.28 μm-, 8.0 ± 0.80 μm-, 10.5 ± 0.75 μm-, and 11.9 ± 0.12 μm-diameter polystyrene (PS) beads with a separation purity of 95%, 92%, 50%, and 63%, respectively. The 10.5 μm- and 11.9 μm-diameter PS beads have relatively low separation purity of 50% and 63%. However, at an applied voltage of 8 Vrms, we separate 11.9 μm-diameter PS beads with a separation purity over 99%. At an applied voltage of 500 kHz, 10 Vrms sinusoidal wave and a flow rate of 0.11 μl min−1, we separate red blood cells (5.4 ± 1.3 μm-diameter) and white blood cells (8.1 ± 1.5 μm-diameter) with a separation purity over 99%. Therefore, the present particle separator achieves clog-free, size-dependent particle separation, which is capable of size tuning of separable particles.Keywords
This publication has 16 references indexed in Scilit:
- Continuous dielectrophoretic cell separation microfluidic deviceLab on a Chip, 2006
- The hydrodynamic focusing effect inside rectangular microchannelsJournal of Micromechanics and Microengineering, 2006
- Research HighlightsLab on a Chip, 2005
- When Microfluidic Devices Go BadAnalytical Chemistry, 2005
- Blood-on-a-ChipAnnual Review of Biomedical Engineering, 2005
- Continuous particle separation in a microchannel having asymmetrically arranged multiple branchesLab on a Chip, 2005
- Biomimetic technique for adhesion-based collection and separation of cells in a microfluidic channelLab on a Chip, 2004
- Continuous Particle Separation Through Deterministic Lateral DisplacementScience, 2004
- Functional polymer microspheresProgress in Polymer Science, 2000
- Dielectrophoretic separation of cancer cells from bloodIEEE Transactions on Industry Applications, 1997