Histologic and Biomechanical Evaluation of Crosslinked and Non-Crosslinked Biologic Meshes in a Porcine Model of Ventral Incisional Hernia Repair
Authors:
- Corey R. DeekenCDCorey R. Deeken
- Lora MelmanLMLora Melman
- Eric D. JenkinsEJEric D. Jenkins
- Suellen C. GrecoSGSuellen C. Greco
- Margaret M. FrisellaMFMargaret M. Frisella
- Brent D. MatthewsBMBrent D. Matthews
- 1 May 2011
- journal article
- Published by Wolters Kluwer Health in Journal of the American College of Surgeons
- Vol. 212 (5) , 880-888
Abstract
The objective of this study was to evaluate the biomechanical characteristics and histologic remodeling of crosslinked (Peri-Guard, Permacol) and non-crosslinked (AlloDerm, Veritas) biologic meshes over a 12 month period using a porcine model of incisional hernia repair. Bilateral incisional hernias were created in 48 Yucatan minipigs and repaired after 21 days using an underlay technique. Samples were harvested at 1, 6, and 12 months and analyzed for biomechanical and histologic properties. The same biomechanical tests were conducted with de novo (time 0) meshes as well as samples of native abdominal wall. Statistical significance (p < 0.05) was determined using 1-way analysis of variance with a Fisher's least significant difference post-test. All repair sites demonstrated similar tensile strengths at 1, 6, and 12 months and no significant differences were observed between mesh materials (p > 0.05 in all cases). The strength of the native porcine abdominal wall was not augmented by the presence of the mesh at any of the time points, regardless of de novo tensile strength of the mesh. Histologically, non-crosslinked materials showed earlier cell infiltration (p < 0.01), extracellular matrix deposition (p < 0.02), scaffold degradation (p < 0.05), and neovascularization (p < 0.02) compared with crosslinked materials. However, by 12 months, crosslinked materials showed similar results compared with the non-crosslinked materials for many of the features evaluated. The tensile strengths of sites repaired with biologic mesh were not impacted by very high de novo tensile strength/stiffness or mesh-specific variables such as crosslinking. Although crosslinking distinguishes biologic meshes in the short-term for histologic features, such as cellular infiltration and neovascularization, many differences diminish during longer periods of time. Characteristics other than crosslinking, such as tissue type and processing conditions, are likely responsible for these differences.Keywords
This publication has 19 references indexed in Scilit:
- Non–Cross-Linked Porcine Acellular Dermal Matrices for Abdominal Wall ReconstructionPlastic and Reconstructive Surgery, 2010
- Host Response to Implanted Porcine-Derived Biologic Materials in a Primate Model of Abdominal Wall RepairTissue Engineering Part A, 2008
- Host Response to Human Acellular Dermal Matrix Transplantation in a Primate Model of Abdominal Wall RepairTissue Engineering Part A, 2008
- The extracellular matrix as a biologic scaffold materialBiomaterials, 2007
- Host response after reconstruction of abdominal wall defects with porcine dermal collagen in a rat modelAmerican Journal of Obstetrics and Gynecology, 2004
- Long-term Follow-up of a Randomized Controlled Trial of Suture Versus Mesh Repair of Incisional HerniaAnnals of Surgery, 2004
- Xenogeneic extracellular matrix as a scaffold for tissue reconstructionTransplant Immunology, 2004
- Laparoscopic Repair of Ventral HerniasAnnals of Surgery, 2003
- Comparative evaluation of treated bovine pericardium as a xenograft for hernia repairBiomaterials, 1991
- Incisional hernia: A 10 year prospective study of incidence and attitudesBritish Journal of Surgery, 1985