Accuracy of intramedullary alignment in total knee replacement
Authors:
- M.P. ManningMMM.P. Manning
- 1 September 1992
- journal article
- Published by IOP Publishing in Journal of Biomedical Engineering
- Vol. 14 (5) , 363-370
Abstract
The accuracy of a system of intramedullary alignment using 6 mm rods was assessed in 100 patients undergoing total knee replacements. Post-operative, full length weight-bearing X-rays were used; the mechanical axis from head was used as the reference axis. The method of calculating the errors produced by flexion and rotation of the limb in relation to the X-ray beam is described, the mean deviation from the mechanical axis in 100 cases being 0.67 degrees valgus with a standard deviation of 2.47 degrees. The maximum error was 6.68 degrees valgus and 4.62 degrees varus. The purpose of this study is twofold, first to assess the accuracy of this system of intramedullary alignment and, second, to develop a method of correcting for apparent radiological misalignment using standard radiographic equipment.Keywords
This publication has 6 references indexed in Scilit:
- Tibiofemoral alignment and the results of knee replacementThe Journal of Bone and Joint Surgery. British volume, 1985
- Alignment of Total Knee ComponentsOrthopedics, 1984
- Alignment in Total Knee ArthroplastyClinical Orthopaedics and Related Research, 1983
- Stress fractures after total knee arthroplasty.Journal of Bone and Joint Surgery, 1980
- Mechanics of the knee and problems in reconstructive surgeryThe Journal of Bone and Joint Surgery. British volume, 1978
- Influence of positioning of prosthesis in total knee replacementJournal of Bone and Joint Surgery, 1977