A Framework for Measuring the Time-Varying Shape and Full-Field Deformation of Residual Limbs Using 3-D Digital Image Correlation
- 24 January 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 66 (10) , 2740-2752
Abstract
Effective prosthetic socket design following lower-limb amputation depends upon the accurate characterization of the shape of the residual limb as well as its volume and shape fluctuations. Objective: This study proposes a novel framework for the measurement and analysis of residual limb shape and deformation, using a high-resolution and low-cost system. Methods: A multi-camera system was designed to capture sets of simultaneous images of the entire residuum surface. The images were analyzed using a specially developed open-source three-dimensional digital image correlation (3D-DIC) toolbox, to obtain the accurate time-varying shapes as well as the full-field deformation and strain maps on the residuum skin surface. Measurements on a transtibial amputee residuum were obtained during knee flexions, muscle contractions, and swelling upon socket removal. Results: It was demonstrated that 3D-DIC can be employed to quantify with high resolution the time-varying residuum shapes, deformations, and strains. Additionally, the enclosed volumes and cross-sectional areas were computed and analyzed. Conclusion: This novel low-cost framework provides a promising solution for the in-vivo evaluation of residuum shapes and strains, as well as the potential for characterizing the mechanical properties of the underlying soft-tissues. Significance: These data may be used to inform data-driven computational algorithms for the design of prosthetic sockets, as well as of other wearable technologies mechanically interfacing with the skin.Keywords
All Related Versions
Funding Information
- National Institutes of Health (R01EB024531)
This publication has 40 references indexed in Scilit:
- An Investigation of Comfort Level Trend Differences Between the Hands-On Patellar Tendon Bearing and Hands-Off Hydrocast Transtibial Prosthetic SocketsJPO Journal of Prosthetics and Orthotics, 2011
- Modeling the Mechanical Response of In Vivo Human Skin Under a Rich Set of DeformationsAnnals of Biomedical Engineering, 2011
- Measuring the mechanical properties of human skin in vivo using digital image correlation and finite element modellingThe Journal of Strain Analysis for Engineering Design, 2009
- Digital image correlation and finite element modelling as a method to determine mechanical properties of human soft tissue in vivoJournal of Biomechanics, 2009
- Review of secondary physical conditions associated with lower-limb amputation and long-term prosthesis useJournal of Rehabilitation Research and Development, 2008
- Internal mechanical conditions in the soft tissues of a residual limb of a trans-tibial amputeeJournal of Biomechanics, 2008
- A method for aligning trans-tibial residual limb shapes so as to identify regions of shape changeIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2005
- Use and satisfaction with prosthetic limb devices and related services 11No commercial party having a direct financial interest in the results of the research supporting this article has or will confer a benefit on the author(s) or on any organization with which the author(s) is/are associated.Archives of Physical Medicine and Rehabilitation, 2004
- Conventional and Hydrostatic Transtibial Interface ComparisonJPO Journal of Prosthetics and Orthotics, 1999
- Relaxed Skin Tension Lines (RSTL) versus Other Skin LinesPlastic and Reconstructive Surgery, 1984