Abstract
Comprehensive verification of the intricate dose distributions associated with advanced radiation treatments is now an immediate and substantial problem. The task is challenging using traditional dosimeters because of restrictions to point measurements (ion chambers, diodes, TLD, etc.) or planar measurements (film). In essence, rapid advances in the technology to deliver radiation treatments have not been paralleled by corresponding advances in the ability to verify these treatments. A potential solution has emerged in the form of water equivalent three dimensional (3D) gel‐dosimetry. In this paper we present basic characterization and performance studies of a prototype optical‐CT scanning system developed in our laboratory. An analysis of the potential role or scope of gel dosimetry, in relation to other dosimeters, and to verification across the spectrum of therapeutic techniques is also given. The characterization studies enabled the determination of nominal operating conditions for optical‐CT scanning. “Finger” phantoms are introduced as a powerful and flexible tool for the investigation of optical‐CT performance. The modulation‐transfer function (MTF) of the system is determined to be better than 10% out to 1 mm−1, confirming sub‐mm imaging ability. System performance is demonstrated by the acquisition of a 1×1×1 mm3 dataset through the dose distribution delivered by an x‐ray lens that focuses x rays in the energy range 40–80 KeV. This 3D measurement would be extremely difficult to achieve with other dosimetry techniques and highlights some of the strengths of gel dosimetry. Finally, an optical Monte Carlo model is introduced and shown to have potential to model light transport through gel‐dosimetry systems, and to provide a tool for the study and optimization of optical‐CT gel dosimetry. The model utilizes Mie scattering theory and requires knowledge of the variation of the particle size distribution with dose. The latter was determined here using the technique of dynamic‐light‐scattering.Keywords
This publication has 14 references indexed in Scilit:
- Intensity-Modulated RadiotherapyThe Cancer Journal, 2002
- Artefacts in multi-echoT2imaging for high-precision gel dosimetry: III. Effects of temperature drift during scanningPhysics in Medicine & Biology, 2001
- High resolution gel‐dosimetry by optical‐CT and MR scanningMedical Physics, 2001
- New trends in prostatic cancer research. Three-dimensional conformal radiation therapy (3-D CRT), brachytherapy, and new therapeutic modalities.2001
- Experimental determination of the diffusion coefficient in two-dimensions in ferrous sulphate gels using the finite element methodAustralasian Physics & Engineering Sciences in Medicine, 2001
- Artefacts in multi-echoT2imaging for high-precision gel dosimetry: II. Analysis ofB1-field inhomogeneityPhysics in Medicine & Biology, 2000
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998
- Optical CT reconstruction of 3D dose distributions using the ferrous–benzoic–xylenol (FBX) gel dosimeterMedical Physics, 1998
- Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: II. Optical properties of the BANG polymer gelPhysics in Medicine & Biology, 1996
- NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: A new approach to 3D dosimetry by MRIMagnetic Resonance Imaging, 1993