Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific oscillometric units in cats
Authors:
- Mark J. AciernoMAMark J. Acierno
- Diana SeatonDSDiana Seaton
- Mark A. MitchellMMMark A. Mitchell
- Anderson da CunhaACAnderson da Cunha
- 15 August 2010
- journal article
- Published by American Veterinary Medical Association (AVMA) in Journal of the American Veterinary Medical Association
- Vol. 237 (4) , 402-406
Abstract
Objective—To determine whether veterinary-specific oscillometric blood pressure units yield measurements that are in good agreement with directly measured blood pressures in cats. Design—Evaluation study. Animals—21 cats undergoing routine spaying or neutering. Procedures—A 24-gauge catheter was inserted in a dorsal pedal artery, and systolic, diastolic, and mean arterial pressures were directly measured with a validated pressure measurement system. Values were compared with indirect blood pressure measurements obtained with 3 veterinary-specific oscillometric blood pressure units. Results—There was poor agreement between indirectly and directly measured blood pressures. For unit 1, bias between indirectly and directly measured values was −14.9 mm Hg (95% limits of agreement [LOA], −52.2 to 22.4 mm Hg), 4.4 mm Hg (95% LOA, −26.0 to 34.8 mm Hg), and −1.3 mm Hg (95% LOA, −26.7 to 24.1 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 2, bias was −10.3 mm Hg (95% LOA, −52.9 to 32.2 mm Hg), 13.0 mm Hg (95% LOA, −32.1 to 58.0 mm Hg), and 9.1 mm Hg (95% LOA, −32.9 to 51.2 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 3, bias was −13.4 mm Hg (95% LOA, −51.8 to 25.1 mm Hg), 8.0 mm Hg (95% LOA, −25.5 to 41.6 mm Hg), and −3.6 mm Hg (95% LOA, −31.6 to 24.5 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. Conclusions and Clinical Relevance—Results suggested that none of the 3 veterinary-specific oscillometric blood pressure units could be recommended for indirect measurement of blood pressure in cats.Keywords
This publication has 11 references indexed in Scilit:
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- Hypertension in renal disease: Diagnosis and treatmentClinical Techniques in Small Animal Practice, 2005
- A Comparison of Indirect Blood Pressure Monitoring Techniques in the Anesthetized CatVeterinary Surgery, 1998
- Evaluation of an Oscillometric Blood Pressure Monitor on Anesthetized Cats and the Effect of Cuff Placement and Fur on AccuracyVeterinary Surgery, 1997
- Doppler Ultrasonographic, Oscillometric Sphygmomanometric, and Photoplethysmographic Techniques for Noninvasive Blood Pressure Measurement in Anesthetized CatsJournal of Veterinary Internal Medicine, 1995
- The Fast Flush Test Measures the Dynamic Response of the Entire Blood Pressure Monitoring SystemAnesthesiology, 1992
- Evaluation of the Doppler ultrasonic method of measuring systolic arterial blood pressure in catsAmerican Journal of Veterinary Research, 1992
- Direct Blood Pressure Measurement —Dynamic Response RequirementsAnesthesiology, 1981
- The Meaning of the Point of Maximum Oscillations in Cuff Pressure in the Indirect Measurement of Blood Pressure—Part IIJournal of Biomechanical Engineering, 1980
- Noninvasive automatic determination of mean arterial pressureMedical & Biological Engineering & Computing, 1979