The Study on Hemodynamic Effect of Varied Support Models of BJUT-II VAD on Coronary Artery
Authors:
- 1 November 2014
- journal article
- adult circulatory-support
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 60 (6) , 643-651
Abstract
BJUT-II VAD (Beijing University of Technology ventricular assist device II) is a novel left ventricular assist device. Because of the special connection between the pump and native heart, the hemodynamic effects of BJUT-II VAD on coronary artery are still unclear. Hence, numerical simulations have been conducted to clarify changes in hemodynamic effects of different support modes. A patient-specific left coronary arterial geometric model is reconstructed based on the computed tomography (CT) data. Three support modes, “constant speed mode,” “co-pulse mode,” and “counter pulse mode,” are used in this study. The wall shear stress (WSS), wall shear stress gradient (WSSG), cycle-averaged wall shear stress (avWSS), oscillatory shear index (OSI), and the flow pattern are calculated to evaluate the hemodynamic states of coronary artery. The computational results demonstrate that the hemodynamic states of coronary artery are directly affected by the support modes. The co-pulse modes could achieve the highest blood perfusion (constant speed: 153 ml/min vs. co-pulse: 775 ml/min vs. counter pulse: 140 ml/min) and the highest avWSS (constant speed: 18.1 Pa vs. co-pulse: 42.6 Pa vs. counter pulse: 22.6 Pa). In addition, both the WSS and WSSG at the time of peak blood velocity under the constant speed mode are lower than those under other two support modes. In contrast, the counter pulse mode generates the highest OSI value (constant speed: 0.365 vs. co-pulse: 0.379 vs. counter pulse: 0.426). BJUT-II VAD under co-pulse mode may have benefits for improving coronary perfusion and preventing the development of atherosclerosis; however, the constant speed mode may have benefit for preventing the development of plaque vulnerability.Keywords
This publication has 31 references indexed in Scilit:
- Comparison of Hemodynamics in the Ascending Aorta Between Pulsatile and Continuous Flow Left Ventricular Assist Devices Using Computational Fluid Dynamics Based on Computed Tomography ImagesArtificial Organs, 2013
- Computational fluid dynamics analysis of surgical adjustment of left ventricular assist device implantation to minimise stroke riskComputer Methods in Biomechanics and Biomedical Engineering, 2013
- Orthotopic heart transplant versus left ventricular assist device: A national comparison of cost and survivalThe Journal of Thoracic and Cardiovascular Surgery, 2013
- Influence of LVAD Cannula Outflow Tract Location on Hemodynamics in the Ascending AortaASAIO Journal, 2012
- Importance of realistic LVAD profiles for assisted aortic simulations: evaluation of optimal outflow anastomosis locationsComputer Methods in Biomechanics and Biomedical Engineering, 2012
- Physiological Control of Intraaorta Pump Based on Heart RateASAIO Journal, 2011
- Modeling and Identification of an Intra-Aorta PumpASAIO Journal, 2010
- A Global Sliding Mode Controller Design for an Intra-Aorta PumpASAIO Journal, 2010
- Phasic coronary blood flow pattern during a continuous flow left ventricular assist supportEuropean Journal of Cardio-Thoracic Surgery, 2005
- Myocardial Hemodynamics, Physiology, and Perfusion With an Axial Flow Left Ventricular Assist Device in the CalfASAIO Journal, 2004