Effect of expandable full-size intraocular lenses on lens centration and capsule opacification in rabbits
Authors:
- Michael BlumenthalMBMichael Blumenthal
- David J. AppleDADavid J. Apple
- 1 March 1999
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 25 (3) , 347-356
Abstract
To study the effects of expandable, hydrogel, full-size lenses (FSLs) on lens centration and posterior capsule opacification (PCO) in the rabbit model. Center for intraocular Lens Research, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA. Small FSLs designed for human eyes were implanted unilaterally in 20 rabbit eyes. Lens centration and PCO in these eyes were compared with those in eyes with conventional poly(methyl methacrylate) (PMMA) intraocular lenses (IOLs) or aphakia in the fellow eyes. In a second experiment, medium-size or large FSLs designed for rabbit eyes were implanted in 20 rabbit eyes and the results compared with those in eyes with conventional PMMA IOLs. Slitlamp examination of all eyes was performed regularly, after enucleation, gross inspection and microscopic studies were performed. Decentration of FSLs occurred in the presence of an anterior capsule tear or very large capsulorhexis. Posterior capsule opacification developed only at sites where the implanted lens was not in contact with the capsule. Aphakic eyes and looped IOLs developed Soemmering's rings peripherally. In eyes with small FSLs, PCO developed between the lens and the center of the posterior capsule, whereas in eyes with large FSLs, only small amounts of central PCO developed. Retained high-viscosity viscoelastic material had no effect on PCO development. Complete refilling of the capsular bag by a full-size IOL was associated with excellent centration and only a small degree of PCO. Elastic lenses may thus serve as a starting point for the development of an artificial lens with accommodative function.Keywords
This publication has 20 references indexed in Scilit:
- The Capsular Bag after Short- and Long-term Fixation of Intraocular LensesOphthalmology, 1995
- Posterior capsule opacificationSurvey of Ophthalmology, 1992
- Side-View Analysis of the LensArchives of Ophthalmology (1950), 1992
- Side-View Analysis of the LensArchives of Ophthalmology (1950), 1992
- Intraocular lens centration with continuous tear capsulotomyJournal of Cataract & Refractive Surgery, 1990
- Posterior capsular opacification and intraocular lens decentration Part I: Comparison of various posterior chamber lens designs implanted in the rabbit modelJournal of Cataract & Refractive Surgery, 1988
- Decentration of Flexible Loop Posterior Chamber Intraocular Lenses in a Series of 222 Postmortem EyesOphthalmology, 1988
- Accommodation of an Endocapsular Silicone Lens (Phaco-Ersatz ) in the Nonhuman PrimateOphthalmology, 1987
- Phaco-Ersatz: Cataract surgery designed to preserve accommodationGraefe's Archive for Clinical and Experimental Ophthalmology, 1986