Pseudobacteremia due to contaminated alcohol swabs
Author:
- S A BergerSBS A Berger
- 1 October 1983
- journal article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 18 (4) , 974-5
Abstract
Fifteen isolates of Bacillus species were recovered during a 19-day period after the introduction of an automated radiometric blood culture system. The outbreak was traced to alcohol-resistant spores present in the cotton used to disinfect blood culture bottles. The role of Bacillus species as a pathogen and a laboratory contaminant is discussed.Keywords
This publication has 7 references indexed in Scilit: