Determination of Human Apolipoproteins A-I, B, and E by Laser Nephelometry

Authors:
  • PWP. Weisweiler
    P. Weisweiler
  • PSP. Schwandt
    P. Schwandt

Abstract
A laser immunonephelometric procedure for the determination of human apolipoproteins A-I, B and E in serum and lipoproteins was developed. Coefficients of variation were 2.3-3.4% within-run, 4.7-4.9% between-run. Serum apolipoprotein A-I values were similar in normal and hyperlipoproteinemic subjects with a correlation coefficient of r = 0.585 to HDL [high density lipoprotein]-cholesterol levels. LDL[low density lipoprotein]-apolipoprotein B values were highly (r = 0.935), VLDL [very low density lipoprotein]-apolipoprotein B values less strongly (r = 0.472) correlated to the corresponding cholesterol levels. Serum apolipoprotein E concentrations were distinctly elevated in type III and not correlated to cholesterol values.

This publication has 13 references indexed in Scilit: