Microstructural andin vitro chemical investigations into plasma-sprayed bioceramic coatings
- 1 January 1998
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 43 (4) , 441-450
Abstract
Hydroxyapatite (HA) coatings plasma sprayed without and with bond coats (titania, zirconia) onto titanium alloy (Ti6A14V) substrates under both atmospheric and low pressure plasma spray conditions were investigated in terms of their microstructure and their resorption resistance during immersion in simulated body fluid (Hank's balanced salt solution). The microstructures of test samples were characterized using SEM on as-sprayed and leached surfaces and on the corresponding cross sections. Selected coating systems were studied by 2-dimensional secondary ion mass spectroscopy imaging to obtain information on plasma spray induced diffusional processes at the coating interfaces, as well as the spatial distribution of minor and trace elements. Coatings consisting of thin (10–15 μm) titania/zirconia (eutectic ratio) and titania bond coats, combined with a 150- to 180-μm thick HA top coat, yielded peel strengths in excess of 32 N/m, as well as sufficient resorption resistance. © 1998 John Wiley & Sons, Inc. J Biomed Mater Res (Appl Biomater) 43: 441–450, 1998Keywords
This publication has 15 references indexed in Scilit:
- Acceleration and Deceleration of Bone-Like Crystal Growth on Ceramic Hydroxyapatite by Electric PolingChemistry of Materials, 1996
- Formation of hydroxyapatite coating on pure titanium substrates by ion beam dynamic mixingSurface and Coatings Technology, 1994
- Chemistry and properties of solids with the [NZP] skeletonSolid State Ionics, 1993
- Hydroxyapatite ceramic coating for bone implant fixationActa Orthopaedica, 1993
- Hydroxyapatite coating modifies implant membrane formationActa Orthopaedica, 1992
- Quality assurance of joint replacement Legal regulation and medical judgementActa Orthopaedica, 1992
- Plasma‐sprayed coatings of tetracalciumphosphate, hydroxyl‐apatite, and α‐TCP on titanium alloy: An interface studyJournal of Biomedical Materials Research, 1991
- Plasma sprayed coatings of hydroxylapatiteJournal of Biomedical Materials Research, 1987
- Use of tricalcium phosphate or electrical stimulation to enhance the bone–porous implant interfaceJournal of Biomedical Materials Research, 1986
- Fixation of porous titanium implants in cortical bone enhanced by electrical stimulationJournal of Biomedical Materials Research, 1981