Abstract
Characteristic x‐ray production with energetic electrons depends strongly on the overvoltage, the ratio of the incident beam energy to the critical excitation energy for the atomic species of interest. Low‐voltage x‐ray micro‐analysis (beam energy ≤ 5 keV) is especially susceptible to artifacts due to sample charging because the overvoltage is low and even slight charging can strongly affect peak intensities. The Duane‐Hunt bremsstrahlung limit is a good diagnostic to detect sample charging. Dynamic charging effects, however, can influence spectra despite an apparently satisfactory Duane‐Hunt limit. Dynamic charging effects must be examined by time series experiments, or through use of dynamic energy windows continuously measuring count rates placed across the spectrum. When charging is a problem, conductive surface coatings can eliminate the effects. When pristine surfaces must be examined without coating, the use of a conductive grid can control charging so that useful x‐ray spectra can be obtained.