The effect of patient positioning on measurements of bone mineral density of the proximal femur: a simulation study using computed tomographic images
Authors:
- 24 February 2023
- journal article
- research article
- Published by Springer Nature in Archives of Osteoporosis
- Vol. 18 (1) , 1-11
Abstract
Summary The patient’s position may affect the bone mineral density (BMD) measurements; however, the extent of this effect is undefined. This CT image-based simulation study quantified changes in BMD induced by hip flexion, adduction, and rotations to recommend appropriate patient positioning when acquiring dual-energy x-ray absorptiometry images. Purpose Several studies have analyzed the effect of hip rotation on the measurement of bone mineral density (BMD) of the proximal femur by dual-energy x-ray absorptiometry (DXA). However, as the effects of hip flexion and abduction on BMD measurements remain uncertain, a computational simulation study using CT images was performed in this study. Methods Hip CT images of 120 patients (33 men and 87 women; mean age, 82.1 ± 9.4 years) were used for analysis. Digitally reconstructed radiographs of the proximal femur region were generated from CT images to calculate the BMD of the proximal femur region. BMD at the neutral position was quantified, and the percent changes in BMD when hip internal rotation was altered from −30° to 15°, when hip flexion was altered from 0° to 30°, and when hip abduction was altered from –15° to 30° were quantified. Analyses were automatically performed with a 1° increment in each direction using computer programming. Results The alteration of hip angles in each direction affected BMD measurements, with the largest changes found for hip flexion (maximum change of 17.7% at 30° flexion) and the smallest changes found for hip rotation (maximum change of 2.2% at 15° internal rotation). The BMD measurements increased by 0.34% for each 1° of hip abduction, and the maximum change was 12.3% at 30° abduction. Conclusion This simulation study quantified the amount of BMD change induced by altering the hip position. Based on these results, we recommend that patients be positioned carefully when acquiring DXA images.Keywords
This publication has 28 references indexed in Scilit:
- Best Practices for Dual-Energy X-ray Absorptiometry Measurement and Reporting: International Society for Clinical Densitometry GuidanceJournal of Clinical Densitometry, 2016
- Difference in Stem Alignment Between the Direct Anterior Approach and the Posterolateral Approach in Total Hip ArthroplastyThe Journal of Arthroplasty, 2015
- Changes in Femoral Version During Implantation of Anatomic Stems: Implications on Stem DesignThe Journal of Arthroplasty, 2015
- Clinical Use of Quantitative Computed Tomography (QCT) of the Hip in the Management of Osteoporosis in Adults: the 2015 ISCD Official Positions—Part IJournal of Clinical Densitometry, 2015
- Diagnostic criteria for primary osteoporosis: year 2012 revisionJournal of Bone and Mineral Metabolism, 2013
- Effect of a Supervised Hip Flexor Stretching Program on Gait in Frail Elderly PatientsPM&R, 2011
- Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems?Osteoporosis International, 2009
- Range of change of measured BMD in the femoral neck and total hip with rotation in womenJournal of Bone and Mineral Metabolism, 2004
- Effect of Leg Rotation on Hip Bone Mineral Density MeasurementsJournal of Clinical Densitometry, 2003
- Effect of femoral rotation on bone mineral density measurements with dual energy X-ray absorptiometryCalcified Tissue International, 1995