Agreement Between Transcutaneous Monitoring and Arterial Blood Gases During COPD Exacerbation
- 1 June 2021
- journal article
- research article
- Published by SAGE Publications in Respiratory Care
- Vol. 66 (10) , 1560-1566
Abstract
BACKGROUND: Transcutaneous measurements of CO 2 and O 2 ( P tcCO 2 , P tcO 2 ) are noninvasive and allow for continuous monitoring in adults with exacerbation of COPD, but substantial accuracy issues may exist. We investigated agreement between results of arterial blood gas analysis and transcutaneous measurements of CO 2 and O 2 in patients with COPD. METHODS: Adult subjects were monitored after acute admission to a respiratory intermediate care unit or ICU due to exacerbation of COPD and with ongoing noninvasive ventilation or immediately following extubation. Monitored variables were continuous transcutaneous measurement and simultaneous routine arterial blood gas analysis. Agreement between measurements was assessed by calculating bias with 95% limits of agreement for single-point estimates of P tcCO 2 versus P aCO 2 and versus P aO 2 , and for changes in transcutaneous measurements between 2 time points ( Δ P tcCO 2 and Δ P tcO 2 ). We considered limits of agreement within ± 7.5 mm Hg to be acceptable. RESULTS: A total of 57 transcutaneous measurements were made in 20 subjects for comparison with concurrent arterial blood gas analysis at 36 time points. The bias (limits of agreement) for P tcCO 2 and P tcO 2 was 2.5 mm Hg (–10.6 to 15.6 mm Hg) and 11.2 mm Hg (–28.2 to 50.6 mm Hg), respectively. The bias for Δ P tcCO 2 and Δ P tcO 2 was 2.3 mm Hg (–3.8 to 8.3 mm Hg) and –5.3 mm Hg (–37.5 to 27 mm Hg), respectively. CONCLUSIONS: P tcCO 2 and P tcO 2 did not accurately reflect results from arterial blood gas analyses in this study of mostly hypercapnic subjects. Agreement between changes in CO 2 during the monitoring period was acceptable, however, and transcutaneous monitoring may be used for continuous monitoring of P CO 2 in conjunction with arterial blood gas analysis for reference.Keywords
This publication has 24 references indexed in Scilit:
- Superiority of transcutaneous CO2 over end-tidal CO2 measurement for monitoring respiratory failure in nonintubated patients: A pilot studyJournal of Critical Care, 2016
- ABG needle study: a randomised control study comparing 23G versus 25G needle success and pain scoresEmergency Medicine Journal, 2014
- Techniques for the Measurement and Monitoring of Carbon Dioxide in the BloodAnnals of the American Thoracic Society, 2014
- AARC Clinical Practice Guideline: Transcutaneous Monitoring of Carbon Dioxide and Oxygen: 2012Respiratory Care, 2012
- Assessing PaCO2 in acute respiratory disease: accuracy of a transcutaneous carbon dioxide deviceInternal Medicine Journal, 2011
- Confidence interval estimation for the Bland–Altman limits of agreement with multiple observations per individualStatistical Methods in Medical Research, 2011
- Agreement between arterial and transcutaneous PCO2 in patients undergoing non-invasive ventilationRespiratory Medicine, 2011
- Transcutaneous P co 2 Monitoring During Initiation of Noninvasive VentilationChest, 2007
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- Non-Invasive (Transcutaneous) Monitoring of PCO2 (TcPCO2) in Older AdultsGerontology, 2005