Toward a Servoregulation Controller to Automate CO2 Removal in Wearable Artificial Lungs
- 6 August 2021
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 68 (5) , 698-706
Abstract
A laptop-driven, benchtop control system that automatically adjusts carbon dioxide (CO2) removal in artificial lungs (ALs) is described. The proportional-integral-derivative (PID) feedback controller modulates pump-driven air sweep gas flow through an AL to achieve a desired exhaust gas CO2 partial pressure (EGCO2). When EGCO2 increases, the servoregulator automatically and rapidly increases sweep flow to remove more CO2. If EGCO2 decreases, the sweep flow decreases to reduce CO2 removal. System operation was tested for 6 hours in vitro using bovine blood and in vivo in three proof-of-concept sheep experiments. In all studies, the controller automatically adjusted the sweep gas flow to rapidly (2 level when challenged with changes in inlet blood or target EGCO2 levels. CO2 removal increased or decreased as a function of arterial pCO2 (PaCO2). Such a system may serve as a controller in wearable AL systems that allow for large changes in patient activity or disease status.Keywords
This publication has 25 references indexed in Scilit:
- Removing extra CO2 in COPD patientsCurrent Respiratory Care Reports, 2013
- Thirty-Day In-Vivo Performance of a Wearable Artificial Pump-Lung for Ambulatory Respiratory SupportThe Annals of Thoracic Surgery, 2012
- Ambulatory veno-venous extracorporeal membrane oxygenation: Innovation and pitfallsThe Journal of Thoracic and Cardiovascular Surgery, 2011
- Veno-venous extracorporeal membrane oxygenation with interatrial shunting: A novel approach to lung transplantation for patients in right ventricular failureThe Journal of Thoracic and Cardiovascular Surgery, 2011
- Bridge to lung transplantation using short-term ambulatory extracorporeal membrane oxygenationThe Journal of Thoracic and Cardiovascular Surgery, 2010
- Extracorporeal Membrane Oxygenation for 2009 Influenza A(H1N1) Acute Respiratory Distress SyndromeJAMA, 2009
- Automatic Control of the Extra-Corporal Bypass: System Analysis, Modelling and Evaluation of Different Control Modes / Automatische Regelung der extrakorporalen Zirkulation: Analyse, Modelbildung und Evaluation von RegelstrategienBiomedical Engineering / Biomedizinische Technik, 2004
- Auf dem Weg zum Herz-Lungenmaschinen-Automat: Ein Regelsystem für die Sauerstoffspannung im OxygenatorBiomedical Engineering / Biomedizinische Technik, 1997
- Development of a digital adaptive control system for P O 2 regulation in a membrane oxygenatorJournal of Biomedical Engineering, 1992
- Errors in measurements of CO2 with the use of drying agentsClinica Chimica Acta, 1986