Absolute and Relative Stability of an Optical Frequency Reference Based on Spectral Hole Burning inEu3+:Y2SiO5

Authors:

Abstract
We present and analyze four frequency measurements designed to characterize the performance of an optical frequency reference based on spectral hole burning in ${\mathrm{Eu}}^{3+}\mathrm{\text{:}}{\mathrm{Y}}_{2}{\mathrm{SiO}}_{5}$. The first frequency comparison, between a single unperturbed spectral hole and a hydrogen maser, demonstrates a fractional frequency drift rate of $5\ifmmode\times\else\texttimes\fi{}{10}^{-18}\text{ }\text{ }{\mathrm{s}}^{-1}$. Optical frequency comparisons between a pattern of spectral holes, a Fabry-P\'erot cavity, and an ${\mathrm{Al}}^{+}$ optical atomic clock show a short-term fractional frequency stability of $1\ifmmode\times\else\texttimes\fi{}{10}^{-15}{$\tau${}}^{-1/2}$ that averages down to ${2.5}_{-0.5}^{+1.1}\ifmmode\times\else\texttimes\fi{}{10}^{-16}$ at $$\tau${}=540\text{ }\text{ }\mathrm{s}$ (with linear frequency drift removed). Finally, spectral-hole patterns in two different ${\mathrm{Eu}}^{3+}\mathrm{\text{:}}{\mathrm{Y}}_{2}{\mathrm{SiO}}_{5}$ crystals located in the same cryogenic vessel are compared, yielding a short-term stability of $7\ifmmode\times\else\texttimes\fi{}{10}^{-16}{$\tau${}}^{-1/2}$ that averages down to ${5.5}_{-0.9}^{+1.8}\ifmmode\times\else\texttimes\fi{}{10}^{-17}$ at $$\tau${}=204\text{ }\text{ }\mathrm{s}$ (with quadratic frequency drift removed).

This publication has 31 references indexed in Scilit: