The Use of a Robot to Insert an Electrode Array of Cochlear Implants in the Cochlea: A Feasibility Study and Preliminary Results
Authors:
- 26 April 2021
- journal article
- research article
- Published by S. Karger AG in Audiology and Neurotology
- Vol. 26 (5) , 361-367
Abstract
Introduction: Cochlear implants (CIs) are commonly used for the rehabilitation of profound bilateral hearing loss. However, patients with substantial residual acoustic hearing are potential CI candidates. Because of both improvements in technology and advancements in surgical techniques, it may be possible to preserve hearing to some extent. For more than a decade, it has been suggested that robots are used to perform middle ear surgery. We evaluated the use of the RobOtol® otologic robot specifically to insert CI electrodes into the inner ear. Methods: CI surgery with the conventional approach was performed under general anesthesia. The MED-El Flex 24-electrode array was inserted using RobOtol®. Video recordings were used to calculate the speed of insertion. The positions of the electrodes were evaluated using a cone beam CT. All subjects underwent pure-tone audiometry tests before and after surgery, and the pure-tone average (PTA) was calculated from 250 to 4,000 Hz. Results: The robot inserted implants in 5 patients, and complete insertion of the electrode array was achieved. The speed of insertion of the electrode array was 0.88 ± 0.12 mm/s. The mean loss of the PTA for 5 frequencies (250, 500, 1,000, 2,000, and 4,000 Hz) was 13.60 ± 7.70 dB. Only 1 patient showed a loss of the PTA by >20 dB. For these 5 patients, the cone beam CT findings showed that all the electrode arrays were in the tympanic ramp and had a grade of 0. The results were compared with those obtained from a cohort of 17 patients who underwent manual implantation of a MED-El Flex 24-electrode array. Conclusion: To minimize disturbance to the cochlea while atraumatic electrode arrays are inserted, electrodes can be inserted at a constant, slow speed in the inner ear with the assistance of the RobOtol® robot in a normal clinical surgical setting.Keywords
This publication has 16 references indexed in Scilit:
- Intracochlear Fluid Pressure Changes Related to the Insertional Speed of a CI ElectrodeBioMed Research International, 2014
- Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone SpecimensBioMed Research International, 2014
- The Effects of Insertion Speed on Inner Ear Function during Cochlear Implantation: A Comparison StudyAudiology and Neurotology, 2012
- Friction Force Measurement During Cochlear Implant InsertionOtology & Neurotology, 2012
- Design, Kinematic Optimization, and Evaluation of a Teleoperated System for Middle Ear MicrosurgeryThe Scientific World Journal, 2012
- Implications of Minimizing Trauma During Conventional Cochlear ImplantationOtology & Neurotology, 2011
- Impact of the Insertion Speed of Cochlear Implant Electrodes on the Insertion ForcesOtology & Neurotology, 2011
- Evidence for the Expansion of Adult Cochlear Implant CandidacyEar & Hearing, 2010
- An autonomous surgical robot for drilling a cochleostomy: preliminary porcine trialClinical Otolaryngology, 2008
- Comparative Study of Cochlear Damage With Three Perimodiolar Electrode DesignsThe Laryngoscope, 2003