Metallurgical analyses of failed orthopedic implants
Authors:
- 1 March 1968
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 2 (1) , 1-22
Abstract
Several metallic orthopedic implants which failed in service and for which the medical histories are known are comprehensively examined using metallographic techniques and electron microprobe analyses to determine the causes of failure. The results indicate that fatigue is a major cause of failure and that, in all probability, the fatigue is initiated by poor design or poor material.Keywords
This publication has 1 reference indexed in Scilit:
- Crack-growth phenomena in fatigued poly(methylmethacrylate)Journal of Materials Science, 1966