Identification of robust and reproducible CT‐texture metrics using a customized 3D‐printed texture phantom
Open Access
- 12 January 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 22 (2) , 98-107
Abstract
Objective: The objective of this study was to evaluate the robustness and reproducibility of computed tomography‐based texture analysis (CTTA) metrics extracted from CT images of a customized texture phantom built for assessing the association of texture metrics to three‐dimensional (3D) printed progressively increasing textural heterogeneity.Materials and Methods: A custom‐built 3D‐printed texture phantom comprising of six texture patterns was used to evaluate the robustness and reproducibility of a radiomics panel under a variety of routine abdominal imaging protocols. The phantom was scanned on four CT scanners (Philips, Canon, GE, and Siemens) to assess reproducibility. The robustness assessment was conducted by imaging the texture phantom across different CT imaging parameters such as slice thickness, field of view (FOV), tube voltage, and tube current for each scanner. The texture panel comprised of 387 features belonging to 15 subgroups of texture extraction methods (e.g., Gray‐level Co‐occurrence Matrix: GLCM). Twelve unique image settings were tested on all the four scanners (e.g., FOV125). Interclass correlation two‐way mixed with absolute agreement (ICC3) was used to assess the robustness and reproducibility of radiomic features. Linear regression was used to test the association between change in radiomic features and increased texture heterogeneity. Results were summarized in heat maps.Results: A total of 5612 (23.2%) of 24 090 features showed excellent robustness and reproducibility (ICC ≥ 0.9). Intensity, GLCM 3D, and gray‐level run length matrix (GLRLM) 3D features showed best performance. Among imaging variables, changes in slice thickness affected all metrics more intensely compared to other imaging variables in reducing the ICC3. From the analysis of linear trend effect of the CTTA metrics, the top three metrics with high linear correlations across all scanners and scanning settings were from the GLRLM 2D/3D and discrete cosine transform (DCT) texture family.Conclusion: The choice of scanner and imaging protocols affect texture metrics. Furthermore, not all CTTA metrics have a linear association with linearly varying texture patterns.Keywords
This publication has 32 references indexed in Scilit:
- Applications and limitations of radiomicsPhysics in Medicine & Biology, 2016
- A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability ResearchJournal of Chiropractic Medicine, 2016
- Reproducibility of radiomics for deciphering tumor phenotype with imagingScientific Reports, 2016
- Core samples for radiomics features that are insensitive to tumor segmentation: method and pilot study using CT images of hepatocellular carcinomaJournal of Medical Imaging, 2015
- Statistical analysis of lung nodule volume measurements with CT in a large‐scale phantom studyMedical Physics, 2015
- Robust Radiomics Feature Quantification Using Semiautomatic Volumetric SegmentationPLOS ONE, 2014
- A Review of Feature Reduction Techniques in NeuroimagingNeuroinformatics, 2013
- Reliability of Measuring Lesion Volumes in Transient Ischemic Attack and Minor StrokeStroke, 2010
- Comparison of ICC and CCC for assessing agreement for data without and with replicationsComputational Statistics & Data Analysis, 2008
- The reliability of measuring physical characteristics of spiculated masses on mammographyBritish Journal of Radiology, 2006