Increase in early polyethylene wear after sterilization with ethylene oxideRadiostereometric analyses of 201 total hips
Authors:
Open Access
- 1 January 2003
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 74 (5) , 531-541
Abstract
We evaluated polyethylene wear by measuring femoral head penetration in 201 THA (median age 62 (31-81) years, 117 women) extracted from 5 randomized studies aimed to assess various fixation principles. There were 30 cemented all-polyethylene Lubinus cups sterilized by gamma irradiation in a reduced oxygen environment, 65 porous-coated Trilogy cups with liners gamma-sterilized in inert gas. Moreover, 37 cemented cups were sterilized with ethylene oxide (Reflection all-poly) and 69 porous-coated cups had liners sterilized in ethylene oxide (Reflection). 28 mm femoral heads were used in all cups. The patients were followed with repeated radiostereometric measurements (RSA) up to 2 years. The activity level of the patients was evaluated by a questionnaire. After 2 years, cups with polyethylene sterilized in EtO had almost twice the proximal and 3D penetration rates, as compared with gamma-sterilized polyethylene. The penetration did not differ between the gamma-irradiated designs. Using stepwise linear regression analysis, we found that the type of sterilization, age and weight were the most important predictors and that they determined the direction of the proximal penetration rate. Activity score, male gender and proximal migration of the cup had little effect. The accelerated wear observed with the EtO-sterilized polyethylene causes concerns about long-term problems and especially in younger patients.Keywords
This publication has 18 references indexed in Scilit:
- The effect of activity levels of total hip arthroplasty patients on socket penetrationThe Journal of Arthroplasty, 2001
- Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE)Biomaterials, 1999
- Zirconia as a ceramic biomaterialBiomaterials, 1999
- Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problemBiomaterials, 1996
- Impact of gamma sterilization on clinical performance of polyethylene in the kneeThe Journal of Arthroplasty, 1996
- Histology of tissue adjacent to an HAC-coated femoral prosthesis. A case reportThe Journal of Bone and Joint Surgery. British volume, 1994
- Activity and socket wear in the Charnley low-friction arthroplastyThe Journal of Arthroplasty, 1994
- Osteolysis from a press-fit hydroxyapatite-coated implantThe Journal of Arthroplasty, 1993
- Long-term results of staple capsulorrhaphy for anterior instability of the shoulder.Journal of Bone and Joint Surgery, 1993
- Property changes of UHMW polyethylene hip cup endoprostheses during implantationJournal of Biomedical Materials Research, 1984