Beam profile studies in an x‐ray fluorescence spectrometer
Authors:
- 1 December 1991
- journal article
- research article
- Published by Wiley in X-Ray Spectrometry
- Vol. 20 (6) , 337-341
Abstract
For small samples, the beam intensity in a secondary target conventional x‐ray fluorescence (XRF) spectrometer is often not sufficient. As a step in our efforts to optimize the three‐axial XRF spectrometer in Gothenburg, the intensity profile of the x‐ray beam from the secondary target was studied. The intensity profile was found to be inhomogeneous. For the beam profile studies, mechanical systems were built which can also be used for position determination of small samples. A movable platform for the detector was constructed and equipped with driving and controlling electronics. To improve further the flexibility of the spectrometer, the sample holder was rebuilt to allow for movements of the sample in the x ‐, y ‐ and z ‐directions. The possibility of tilting the sample and thus varying the angle between the incoming beam and the sample plane from about 0 to 90° was included in the mechanics of the sample holder. The arrangement can also be used to centre the samples in the spectrometer and to move the sample for analysis of different parts of the sample.Keywords
This publication has 6 references indexed in Scilit:
- Sample area dependence in quantitative EDXRF analysisX-Ray Spectrometry, 1991
- Measurements of elemental composition in background aerosol on the west coast of SwedenAtmospheric Environment (1967), 1986
- Application of an Energy Dispersive X-Ray Fluorescence Spectrometer to Air Pollution StudiesPhysica Scripta, 1982
- Background reduction of x-ray fluorescence spectra in a secondary target energy dispersive spectrometerNuclear Instruments and Methods, 1979
- Development of a low power monoenergetic X-ray tube for trace element analysisNuclear Instruments and Methods, 1978
- Optimization of a simple spotting procedure for x-ray fluorescence analysis of watersAnalytica Chimica Acta, 1977